![Flight Medical Innovations 60 DL Service Manual Download Page 59](http://html.mh-extra.com/html/flight-medical-innovations/60-dl/60-dl_service-manual_2305321059.webp)
Removing and Reinstalling Modules
Replacing the Oxygen mixer * (iO2 Internal mixer model only)
Flight 60 Service Manual
|
4-29
Replacing the Oxygen mixer
*
(
iO
2
Internal mixer
model only
)
4.15.1 Removing the Oxygen mixer
To remove the oxygen mixer:
1 Remove the detachable battery (see section 4.3) and the internal battery (see
section 4.4).
2. Remove the ventilator cover (see section 4.5).
3. Remove the Power Supply and the power supply base plate + the solenoid (see
section 4.7).
4.
Remove the muffler.
5.
Remove the O
2
leak sensor
6.
Remove the 2 silicon tubes and disconnect the Nebulizer & O2 Mixer cables from
the O
2
mixer Neb board
Figure 53 – Electronic connectors and silicon tubes
7.
Gently pull out the O
2
mixer
Summary of Contents for 60 DL
Page 2: ...TABLE OF CONTENTS FLIGHT 60 Ventilator II ...
Page 3: ......
Page 27: ...Functional Description Pneumatic Diagram Flight 60 Service Manual 3 11 Pneumatic Diagram ...
Page 29: ...Functional Description Electrical Diagram Flight 60 Service Manual 3 13 Electrical Diagram ...
Page 100: ......
Page 110: ......