Safety Instructions
Cautions
Flight 60 Service Manual
|
2-3
Only a FLIGHT MEDICAL approved exhalation valve can be used with the
FLIGHT 60 Ventilator.
Perform a Circuit Test each time a clean circuit/exhalation valve is installed.
This FLIGHT 60 Ventilator has been tested and found to comply with the EMC
limits according to the EN60601-1-1-2 standard class B. These limits are
designed to provide reasonable protection against harmful interference in a
typical medical installation. The equipment generates uses and can radiate
radio frequency energy and, if not installed and used in accordance with these
instructions, may cause harmful interference to other devices in the vicinity.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference with other
devices, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
Reorient or relocate the receiving device.
Increase the distance between the equipment.
Connect the equipment into an outlet on a circuit different from that to which
the device (s) is connected.
Consult the manufacturer for help.
Cautions
Only use medical grade oxygen with the Air/Oxygen Entrainment Mixer or
Oxygen Blending Bag Kit.
Do not place liquid containers in the immediate vicinity or on top of the FLIGHT
60 Ventilator. Liquids that get into the ventilator can cause equipment
malfunction and damage.
After the FLIGHT 60 Ventilator is serviced, it must completely pass an
Operational Verification Procedure (OVP) before being returned to patient use.