Calibration and Verification tests
Operational Verification Procedure (OVP)
7-18
| FLIGHT 60 Service Manual
Circuit test
3. Replace the solenoid and perform the required calibrations (see 7.3 Test table).
7.2.8 Exhalation Valve Leak
To test the exhalation valve:
1. Set the ventilator to the following settings:
Rate
= 10
Ti
= 1.4
ACMV
VCV
= 1400
2. Set the High-Pressure alarm to 99 cmH
2
O.
3. Connect the ventilator to a test lung using a patient circuit. Use the following
settings for the lung:
C
= 20
Rp
= 20
4. Activate ventilation and continue for 15 breaths.
5. Verify that the pressure level is over 72 cmH
2
O. If pressure doesn’t get to 72
cmH
2
O, decrease the compliance or increase the resistance of the test lung. If
pressure becomes too high after changing compliance/resistance, reduce the set
volume from 1400 until achieving the required pressure level.
6. Check for air flow from the underside of the Exhalation Valve during inhalation; If
air flow is noticed there is a leak and the test fails.
If the test fails:
1. Replace the patient circuit. In case of a Flight60 DL replace the exhalation valve
diaphragm.
2. Repeat the following:
Pressure sensors calibration
Pressure relief calibration
Volume factor calibration
Circuit test
3. Replace the solenoid (see 4.19), and perform the required calibrations (see 7.3
Test table)