Removing and Reinstalling Modules
Replacing the Solenoid Assembly
4-34
| FLIGHT 60 Service Manual
3. Connect the silicon tube attached to the female connector to the inner fitting on
the redundancy solenoid.
4. Using the other silicon tube, connect the Solenoid assembly to the outer fitting on
the redundancy solenoid.
Figure 60 – Connecting the Silicon Tubes
Replacing the Solenoid Assembly
4.19.1 Removing the Solenoid Assembly
To remove the solenoid assembly:
1.
Remove the ventilator cover.
2.
For iO
2
internal mixer model - Remove the O
2
leak sensor.
3.
Remove the 2 screws securing the O
2
leak sensor bracket. Remove the bracket.
Figure 61 – Remove O
2
leak sensor bracket