background image

FMS-2530 Tabletop Ultrasonic Scaler V.1 

 

9.3 Flight Handpiece Cable Set / Sterilizable Sleeve of Handpiece Assembly   
The ultrasonic scaler is a multi-frequency unit compatible with all Cavitron

®

 inserts. The system will 

automatically detect the insert without a need to switch settings. 
 
For more oral hygiene care, the handpiece sleeve can be dismantled and autoclaved. See below: 
 
 
 

 
 
 
 

 
 
 

 
 
To take off the handpiece sleeve, please hold the fastening ring, and rotate the sleeve counterclockwise 
from the Flight handpiece cable set assembly. Note: Avoid rotating or twisting the Flight handpiece 
cable.
 
 
9.4 Foot Pedal Controls and Operation 
The Foot Pedal is a two-position momentary switch, which activates both Ultrasonic energy and water flow 
at the insert tip. 
 

 

 

 

 

 

Foot Pedal Released… 
Both ultrasonic activation and 
irrigating flow stop. 

Foot Pedal Depressed Half Way (1

st

 

position) 
The ultrasonic insert is activated 
and irrigating water flow under 
normal function. 

Foot Pedal fully depressed   
(2

nd

position) 

This activates the “Turbo” mode 
increasing extra power. 

 

Section 10: Accessories 

 

1.

 

Ultrasonic insert (optional) 

2.

 

External power supply (AC-AC) 

3.

 

Foot pedal 

4.

 

Flight handpiece cable set 

Sterilizable Sleeve 
for Handpiece 
Enable to dismantle 
and autoclave   

Fastening Ring 
Hold here, rotate handpiece 
sleeve to release. 

Flight Handpiece Cable Set 
Assembly 
Includes handpiece set, tubing 
and fastening ring. 

O-ring 

Summary of Contents for FMS-2530

Page 1: ...ULTRASONIC SCALER FMS 2530 Flight Dental Systems 21 Kenview Blvd Unit 9 Ontario Canada L6T5G7 www flightdentalsystems com Toll Free 1 866 799 0517...

Page 2: ...s 8 Installation Instructions 7 8 1 General Information 8 2 Water Line Requirements 8 3 Electrical Requirements 8 4 Unpacking the System 8 5 System Installation 8 6 Power Connection 8 7 Water Supply L...

Page 3: ...leaving the office The use of an in line water filter is recommended Use the device for the intended purpose only The ultrasonic scaler must be used by qualified and trained personnel in a medical de...

Page 4: ...able parts inside this device Please contact your dealer to inquire about repair service and technical support CAUTION EQUIPMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETIC MIXTURE...

Page 5: ...ATURE 10 C 40 C RELATIVE HUMIDITY 30 75 ATOMOSPHERIC PRESSURE 700 kPa 1060kPa 5 4 Transportation And Storage Environment AMBIENT TEMPERATURE 10 C 70 C RELATIVE HUMIDITY 10 90 ATOMOSPHERIC PRESSURE 500...

Page 6: ...imum operator and patient safety carefully follow section 12 system maintenance and care information detailed in the operating instructions As with high speed handpieces and other dental devices the c...

Page 7: ...FMS 2530 Tabletop Ultrasonic Scaler V 1 7 infection control procedures...

Page 8: ...ystem the dental office water line should be thoroughly flushed prior to connection to the System After flushing the system verify that there are no leaks 8 3 Electrical Requirements Refer to Section...

Page 9: ...System to the insert AC Power Indicator Light Illuminates when the system is switched on Sterilizable Sleeve of Handpiece Holder Securely holds the Hand piece when system is not being used Water flow...

Page 10: ...t Pedal Controls and Operation The Foot Pedal is a two position momentary switch which activates both Ultrasonic energy and water flow at the insert tip Foot Pedal Released Both ultrasonic activation...

Page 11: ...f the handpiece over a sink or drain Check spray temperature to verify fluid is reaching the working end of the insert tip Adjust the water irrigatation control knob to ensure adequate flow for the se...

Page 12: ...nfectant on a clean towel and wipe all surfaces Allow disinfectant solution to air dry Do not spray disinfectant solution directly on the ultrasonic scaler 3 Place a sterilized handpiece sleeve on the...

Page 13: ...acle on the back of the unit 2 Check that the system s plug is fully seated in an appropriate AC receptacle and that AC current is present Power Indicator Light is lit 1 Check that the foot pedal conn...

Page 14: ...pairs during the warranty period are not carried out by an authorized engineer dealer the warranty will expire immediately 16 3 Warranty Exception The warranty stated herein is the sole warranty appli...

Page 15: ...electromagnetic immunity The FMS 2530 is intended for use in the electromagnetic environment specified below The customer or the user of the FMS 2530 should assure that it is used in such an environme...

Page 16: ...environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 V m 150 KHz to 80 MHz 3 V m 80MHz to 2 5 GHz...

Page 17: ...The FMS 2530 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the FMS 2530 can help prevent electromagnetic interfere...

Page 18: ...FMS 2530 Tabletop Ultrasonic Scaler V 1 17...

Reviews: