FMS-2530 Tabletop Ultrasonic Scaler V.1
14
ANNEX I
Guidance and manufacturer’s declaration-electromagnetic emissions
The FMS-2530 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FMS-2530 should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The FMS-2530 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The FMS-2530 is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations
/flicker emissions
IEC 61000-3-3
Compliance
Guidance and manufacturer’s declaration-electromagnetic immunity
The FMS-2530 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FMS-2530 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power
supply
lines
+ 2kV for power
supply
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5 0.5~+ 2kV common
mode
0.5~+ 2kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage
variations on power
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the FMS-2530 requires