FMS-2530 Tabletop Ultrasonic Scaler V.1
12
*Note: Water-based disinfection solutions are
preferred, as some alcohol-based disinfectant
solutions may be harmful and may discolor plastic
materials.
*Note: The electric wire winding (covered by a heat
shrink tube which is protected by the sterilizable
sleeve of the handpiece) is very sensitive to the
disinfectant solution and water. After disinfection,
wipe the surface of the shrink tube with a slightly
damp cloth and dry thoroughly before use.
Section 13: Troubleshooting
Although service and repair of the Ultrasonic Scaler
should be performed by FLIGHT DENTAL SYSTEMS
dealer personnel, the following are some basic
trouble shooting procedures that will help avoid
unnecessary service calls. Generally, check all lines
and connections to and from the system as a loose
plug or connection will often create problems. Check
the settings on the system's knobs.
13.1 Troubleshooting Guide
Problem: System will not operate:
(Power indicator light is not lit.)
1.
Check that the power switch is in the ON position,
and that the AC power cord is fully seated in the
receptacle on the back of the unit
2.
Check that the system's plug is fully seated in an
appropriate AC receptacle, and that AC current is
present.
(Power Indicator Light is lit.)
1.
Check that the foot pedal connector is fully seated
in the foot pedal receptacle on the back of the
system.
2.
Check the foot pedal.
System operates:
(No water flow to insert)
1.
Check that the water control is properly adjusted.
2.
Check that the water supply control valve(s)
(dental office water supply) are open.
(Insert stops vibrating)
1.
Deactivate foot pedal.
2.
Verify insert is in good condition.
3.
Depress foot pedal to try again
Section 14: Disposal of Unit
‧
Original packaging can be kept until the Ultrasonic Scaler is to be disposed of permanently. You can use
it for shipping or storing your Ultrasonic Scaler at any time.
‧
Dispose of the Ultrasonic Scaler in accordance with local and national laws.
Section 15: Disclaimer
FLIGHT DENTAL SYSTEMS considers itself responsible for the effects on safety, reliability and performance
of this product only if:
‧
Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons
authorized by FLIGHT DENTAL SYSTEMS.
‧
The electrical installation of the relevant room complies with the requirements.
The equipment is used in accordance with these instructions for use.