Section 8: Installation Instructions
8.1 General Information
If the installation of your Flight Dental Systems
FMS-2530 ultrasonic scaler is performed by
someone other than a trained Flight Dental
Systems
®
distributor personnel, care should be
taken to observe the following requirements and
recommendations.
8.2 Water Line Requirements
‧
The system's city water supply tube is factory
provided. Do not discard it from the ultrasonic
scaler kit.
‧
Incoming water supply line pressure to the
ultrasonic scaler must be 20 psi (138 kPa
minimum) to 40 psi (276 kPa) maximum. If your
dental water system's supply line pressure is
above 40 psi, install a water pressure regulator
on the water supply line to your ultrasonic scaler.
‧
A manual shut-off valve on the dental water
system supply line should be used so that the
water can be completely shut-off when the office
is unoccupied.
‧
A filter in the dental water system supply line is
recommended so that any particles in the water
supply will be trapped before reaching the
ultrasonic scaler.
‧
After the above installations are completed on
the dental water supply system, the dental office
water line should be thoroughly flushed prior to
connection to the System.
‧
After flushing the system, verify that there are
no leaks.
8.3 Electrical Requirements
Refer to Section 3: Specifications.
8.4 Unpacking the System
Carefully unpack your ultrasonic scaler and
verify that all components and accessories are
included:
1.
Flight Dental Systems FMS-2530 Ultrasonic
scaler main unit with sterilizable sleeve for the
handpiece.
2.
External power supply (AC-AC)
3.
Flight handpiece cable set attached to unit.
4.
Foot pedal.
5.
User manual.
6.
City water supply tubing with inline filter
8.5 System Installation
‧
Place the FMS-2530 unit on a level surface.
‧
Be sure unit is stable and resting on four feet.
‧
Placing unit in direct sunlight may discolor the
plastic housing.
8.6 Power Connection
‧
The ON/OFF control switch and power indicator is
located on the rear panel of the Ultrasonic Scaler.
(See Section 9.1)
‧
Verify the ON/OFF control switch located on the
rear panel of the ultrasonic scaler is in the OFF
position before proceeding.
‧
Plug the power supply AC cord into the back of
the ultrasonic scaler and the AC plug into an
approved outlet.
8.7 Water Supply Line Connection
Connect the free end of the ultrasonic scaler’s
water supply line to the dental water supply line.
Inspect all connections to make certain there are
no leaks.