WaveInjector WI-400
Supplement to operating instruction - ENGLISH
2018-01-19, SUFLUXUS_WIV4-3EN
62
4.2
Installation of chain support, jaw and pipe mounting fixture
• Select the measuring point according to the recommendations in section 4.1.
• For the installation of the transmitter see operating instruction FLUXUS.
• For the connection of the transducers and the power supply to the transmitter see operating instruction FLUXUS.
4.2.1
Pipe preparation
Cleaning the pipe
Good acoustic contact between the transducer and the pipe wall is necessary for the measurement. Therefore, the pipe
surface at the selected measuring point has to be even and clean.
• Clean the whole circumference of the pipe along the length of the WaveInjector.
• Remove any rust, loose paint or scale.
• When cleaning the pipe, make sure not to cause any unevenness on the pipe surface.
Marking the pipe
Before installation, the place is marked where the chain support, jaw or pipe mounting fixtures are mounted:
• For measurements in diagonal arrangement, see Fig. 4.4.
• For measurements in reflection arrangement, see Fig. 4.5.
Caution!
Contact with grinding dust
This may result in injuries (e.g., breathing difficulties, skin reactions, eye irritations).
→
Wear the required personal protective equipment. Observe the applicable rules.
Fig. 4.4:
Steps for the marking (diagonal arrangement)
x
x
x
x
x
x
x
x
x
x
x
x
x
x
step 1
M1
M1
M1
M’
M1
M’
M2
M’
WI Clamp dist.
step 2
step 3
step 4
step 5
M1
Summary of Contents for WaveInjector WI-400 C Series
Page 50: ...WaveInjector WI 400 Erg nzung zur Betriebsanleitung DEUTSCH 2018 01 19 SUFLUXUS_WIV4 3DE 50...
Page 146: ...WaveInjector WI 400 Suppl ment au mode d emploi FRAN AIS 2018 01 19 SUFLUXUS_WIV4 3FR 146...
Page 195: ...WaveInjector WI 400 SUFLUXUS_WIV4 3RU FLUXUS F60x FLUXUS F70x FLUXUS F80x FLUXUS ADM 8027...
Page 208: ...WaveInjector WI 400 2018 01 19 SUFLUXUS_WIV4 3RU 208 1 2 3 4 5...
Page 209: ...WaveInjectorWI 400 209 2018 01 19 SUFLUXUS_WIV4 3RU 6 7 8 9 10...
Page 211: ...WaveInjectorWI 400 211 2018 01 19 SUFLUXUS_WIV4 3RU 4 11 4 12 4 10 1 4 11 4 12 1 2 1 1 2...
Page 220: ...WaveInjector WI 400 2018 01 19 SUFLUXUS_WIV4 3RU 220 4 5 4 36 4 35 4 36 1 2 1 2...
Page 224: ...WaveInjector WI 400 2018 01 19 SUFLUXUS_WIV4 3RU 224 4 3 4 44 WI A 4 44 4 45 4 43 4 44 1 1...
Page 225: ...WaveInjectorWI 400 225 2018 01 19 SUFLUXUS_WIV4 3RU 1 2 l l 4 46 4 47 4 45 1 4 46 1 2 l...
Page 227: ...WaveInjectorWI 400 227 2018 01 19 SUFLUXUS_WIV4 3RU 4 50 4 52 4 50 1 2 2 1 4 51...
Page 229: ...WaveInjectorWI 400 229 2018 01 19 SUFLUXUS_WIV4 3RU 4 54 WI Plate dist 4 55...
Page 233: ...WaveInjectorWI 400 233 2018 01 19 SUFLUXUS_WIV4 3RU 4 63 4 62 1 2 4 63 1 2 2 1 1 2...
Page 241: ...WaveInjectorWI 400 241 2018 01 19 SUFLUXUS_WIV4 3RU A 2 WI WI FLEXIM WI A 2 WI 1 2 3 4 1 2 3 4...