WaveInjector WI-400
Supplement to operating instruction - ENGLISH
2018-01-19, SUFLUXUS_WIV4-3EN
56
3
Product description
Designs
Use of the coupling plate
A coupling plate is mounted between the transducer and the pipe wall. Its cooling effect creates a temperature difference
between the pipe surface and the transducer contact surface.
At the same time, the coupling plate allows a nearly loss-free ultrasonic coupling between the transducer and the pipe
wall.
Acoustic contact is established:
• between the transducer and the coupling plate by means of a plastic coupling foil
• between the coupling plate and the pipe surface by means of a metal coupling foil
Use of the pipe planer
The supplied pipe planer, see Fig. 3.3 is used for:
• the orientation of chain support, jaw and pipe mounting fixture
• the check of evenness of the pipe surface (as a straightedge)
• grading the pipe surface (measuring point)
Fig. 3.1:
WaveInjector WI-400x-xx-C (here: diagonal arrangement)
1 – transducer box with coupling plate
2 – jaw and chain support with chain
Fig. 3.2:
WaveInjector WI-400x-xx-T (here: diagonal arrangement)
1 – transducer box with coupling plate
2 – pipe mounting fixture with threaded rods
Fig. 3.3:
Pipe planer
1
2
1
2
Summary of Contents for WaveInjector WI-400 C Series
Page 50: ...WaveInjector WI 400 Erg nzung zur Betriebsanleitung DEUTSCH 2018 01 19 SUFLUXUS_WIV4 3DE 50...
Page 146: ...WaveInjector WI 400 Suppl ment au mode d emploi FRAN AIS 2018 01 19 SUFLUXUS_WIV4 3FR 146...
Page 195: ...WaveInjector WI 400 SUFLUXUS_WIV4 3RU FLUXUS F60x FLUXUS F70x FLUXUS F80x FLUXUS ADM 8027...
Page 208: ...WaveInjector WI 400 2018 01 19 SUFLUXUS_WIV4 3RU 208 1 2 3 4 5...
Page 209: ...WaveInjectorWI 400 209 2018 01 19 SUFLUXUS_WIV4 3RU 6 7 8 9 10...
Page 211: ...WaveInjectorWI 400 211 2018 01 19 SUFLUXUS_WIV4 3RU 4 11 4 12 4 10 1 4 11 4 12 1 2 1 1 2...
Page 220: ...WaveInjector WI 400 2018 01 19 SUFLUXUS_WIV4 3RU 220 4 5 4 36 4 35 4 36 1 2 1 2...
Page 224: ...WaveInjector WI 400 2018 01 19 SUFLUXUS_WIV4 3RU 224 4 3 4 44 WI A 4 44 4 45 4 43 4 44 1 1...
Page 225: ...WaveInjectorWI 400 225 2018 01 19 SUFLUXUS_WIV4 3RU 1 2 l l 4 46 4 47 4 45 1 4 46 1 2 l...
Page 227: ...WaveInjectorWI 400 227 2018 01 19 SUFLUXUS_WIV4 3RU 4 50 4 52 4 50 1 2 2 1 4 51...
Page 229: ...WaveInjectorWI 400 229 2018 01 19 SUFLUXUS_WIV4 3RU 4 54 WI Plate dist 4 55...
Page 233: ...WaveInjectorWI 400 233 2018 01 19 SUFLUXUS_WIV4 3RU 4 63 4 62 1 2 4 63 1 2 2 1 1 2...
Page 241: ...WaveInjectorWI 400 241 2018 01 19 SUFLUXUS_WIV4 3RU A 2 WI WI FLEXIM WI A 2 WI 1 2 3 4 1 2 3 4...