18 Outputs
FLUXUS ADM 8027, F801, ADM 8127B
UMFLUXUS_F8V4-6-1EN, 2018-10-10
131
18.7.5
Alarm State Indication
The alarm state can be displayed after the configuration of the alarm outputs and during the measurement. This function is
activated in
Special Funct.
\
SYSTEM settings
\
Dialogs/Menus
. The activation of this function is recommended
when alarm outputs have to be reconfigured frequently.
If the alarm state indication is activated, the alarm output state is displayed after configuring the alarm outputs:
18.8
Deactivation of the Outputs
If the programmed outputs are no longer required, they can be deactivated. The configuration of a deactivated output is
stored and will be available if the output is activated again.
Note!
There is no visual or acoustic indication of alarm output switching.
Select the menu item
SHOW RELAIS STAT
. Select
on
to activate the arlarm state indica-
tion.
The alarm state indication is structured as follows:
Rx =
, with x being the number of the alarm output and
a picto-
gram according to Tab. 18.6.
The configuration of the alarm outputs can be repeated by pressing key CLR. If the config-
uration of the alarm outputs is finished, press enter. The main menu will be displayed.
If the alarm output indication is activated, the alarm state can be displayed during the measurement. Press key
to
scroll through the upper line or key
in the lower line until the alarm state is displayed.
Tab. 18.6: Pictograms for the alarm state indication
No.
func
(switching condition)
typ
(holding behavior)
mode
(switching function)
current
state
R
=
1
OFF
NON-HOLD
NO Cont.
CLOSED
2
MAX
HOLD
NC Cont.
OPEN
3
MIN
+
→
- -
→
+
QUANT.
ERROR
Example:
R1 =
Select
no
in
Output Options\Alarm Output
to deactivate an output. Press ENTER.
SHOW RELAIS STAT
off
>ON<
R1=
R2=
R3=
C=REPEAT
Alarm Output
>NO<
yes