14
15
The new SoftTouch Masks feature an adaptive edge made of a
soft biocompatible material ensuring an excellent adherence
to the face and a Dispersion Limiting Vent. These innovations,
which make them stand out from the rest, allow increased
sedimentation of medication in the lungs and reduced disper-
sion into the air.
USE OF NEBULIZATION MANUAL CONTROL
To achieve continuous nebulisation action you should not use the manual nebulisation control (C6), especially in
the case of children or persons with reduced physical, sensory, or mental capabilities. The manual nebulisation
control is useful for limiting dispersion of the medication in the surrounding environment.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
method B:
Sanitise the accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6
by boiling them in water for 10 minutes; use
demineralised or distilled water to prevent calcium deposits.
method C:
Sanitise the accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6
with a hot steam steriliser for baby-bottle
(not the microwave type). Perform the process faithfully following the instructions of the steriliser. To ensure that the
disinfection is effective, choose a steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a paper towel. Alternatively, dry
them with a jet of hot air (for example, from a hair dryer).
CLEANING IN A CLINICAL OR HOSPITAL SETTING - DISINFECTION AND STERILISATION
Before disinfection or sterilisation, sanitise the nebuliser cup and the accessories, choosing one of the methods described
below.
method A:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6
under potable hot water (approximately 40°C)
with a gentle, non abrasive dish detergent.
method B:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6
in the dishwasher with a hot cycle.
DISINFECTION
Accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5-C6
can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlorite) specific for disinfecting,
which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable water and disinfectant,
according to the proportions indicated on the packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air bubbles on the parts. Leave
the parts immersed for the amount of time indicated on the packaging of the disinfectant associated with the
concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
If you want to also perform the STERILISATION, jump to the STERILISATION paragraph.
After having disinfected the accessories, shake them vigorously and place them on a paper towel. Alternatively, dry
them with a jet of hot air (for example, from a hair dryer).
STERILISATION
Accessories
C1 (Assembled nebulizer)-C2-C3-C4-C5
can be sterilised.
Device: Fractionated vacuum overpressure steam steriliser in accordance with EN 13060.
Implementation:
Wrap every single part to be treated with a sterile barrier system or packaging in accordance with Norm
EN 11607. Place the packed components in the steam steriliser. Run the sterilisation cycle according to the operating
instructions of the device by selecting a temperature of 134°C and a time of 10 minutes first.
Storage:
Store the sterilised parts as per the instructions for use of either the sterile barrier system or packaging.
The sterilisation procedure is validated in its conformity to ISO 17665-1.
At the end of each use store the device complete with accessories in a dry place away from dust.
AIR FILTERING
The device is fitted with a suction filter, located inside the pressure-adjustment
knob, which must be replaced when it is dirty or discoloured. To replace it, lift the
pressure-adjustment Knob (A3) and extract the Filter.
Do not wash or reuse the same Filter. Regular filter replacement is necessary to as-
sure proper compressor performance. The Filter must be regularly checked. The
filter was designed to always remain secured in its seat. Do not replace the filter
during use.
Only use original accessories and spare parts by Flaem, we disclaim any liability
in the event of using non original spare parts or accessories.