6WDUWXS
&RQQHFWLQJ3RZHU
)LVKHU6FLHQWLILF
*UDYLW\&RQYHFWLRQDQG)RUFHG$LU,QFXEDWRUV
&RQQHFWLQJWRWKH3RZHU6XSSO\6RXUFH
1. Before connecting the incubator to the power source, check to see if the power supply
voltage corresponds with the specifications on the nameplate on the front of the incubator.
If the voltage (V) and current (A) ratings given are not as required, do not connect the
incubator to the power source!
2. Connect the IEC connector to the socket at the rear of the incubator.
3. Route the power cord along a path that does not cross exhaust air piping or passageways
and aisles.
With stacked devices, keep the power cord away from hot spots on the other incubator in
the stack.
4. Connect the protection-earthed plug of the power cord to a correctly protection-earthed
and earth leakage circuit breaker fused power socket.
5. Make sure the power cord is not subjected to tensile or compressive force.
.HHSWKHSRZHURXWOHWDFFHVVLEOH
7RDOORZDUDSLGGLVFRQQHFWLRQRISRZHULQFDVHRIDQHPHUJHQF\
PDNHVXUHWKDWSRZHURXWOHWVUHPDLQIUHHO\DFFHVVLEOHDWDOOWLPHV
)LJXUH$&3RZHU6XSSO\6RFNHW
1RWH
The alarm contact is not functional with gravity convection incubators. If you have a need for
alarming, please contact Fisher Scientific Customer Support for advice.
Summary of Contents for Isotemp 100L Incbtr Grvity 120V
Page 1: ......
Page 6: ... LVW RI LJXUHV LL UDYLW RQYHFWLRQ DQG RUFHG LU QFXEDWRUV LVKHU 6FLHQWLILF ...
Page 32: ...3URGXFW HVFULSWLRQ 6KHOI 6 VWHP UDYLW RQYHFWLRQ DQG RUFHG LU QFXEDWRUV LVKHU 6FLHQWLILF ...
Page 58: ...6KXW GRZQ 6KXWWLQJ WKH QFXEDWRU RZQ UDYLW RQYHFWLRQ DQG RUFHG LU QFXEDWRUV LVKHU 6FLHQWLILF ...
Page 68: ...0DLQWHQDQFH 5HWXUQV IRU 5HSDLU UDYLW RQYHFWLRQ DQG RUFHG LU QFXEDWRUV LVKHU 6FLHQWLILF ...
Page 74: ... UURU RGHV UDYLW RQYHFWLRQ DQG RUFHG LU QFXEDWRUV LVKHU 6FLHQWLILF ...
Page 84: ... HYLFH RJ UDYLW RQYHFWLRQ DQG RUFHG LU QFXEDWRUV LVKHU 6FLHQWLILF ...
Page 86: ... UDYLW RQYHFWLRQ DQG RUFHG LU QFXEDWRUV LVKHU 6FLHQWLILF ...