Firesafe 829-0301 Instructions For Use Manual Download Page 9

 

4.5. 

During Use 

Continue to monitor the level of gas flow during the procedure or therapy and the contents level of 
the cylinder (if applicable) and be aware that the supply tubing and hose (if applicable) may be a 
trip hazard. 

4.6. 

After Use 

If the supply is from a gas cylinder, close the cylinder valve and disconnect the flowmeter when 
therapy is no longer required. 

 5.

 

Cleaning and Disinfection 

Ensure the flowmeter is disconnected from the gas supply before attempting to clean it. 

Caution! 

The flowmeter is not suitable for autoclaving. 

5.1. 

Cleaning 

Wipe  over  the  outside  of  the  flowmeter  and  the  gas  supply  hose  with  an  alcohol  or  disinfecting 
wipe. 

5.2. 

Suspected Contamination 

If you suspect that the flowmeter is contaminated, remove it from use and refer the device to the 
appropriate department. 

Never immerse the flowmeter in any fluid or attempt to clean internal parts. 

 6.

 

Maintenance 

6.1. 

Interim Inspection 

The flowmeter should be regularly cleaned, inspected for damage and checked for performance.  
The frequency of inspection and performance checks depends upon usage.  As a guideline, if the 
flowmeter is used daily this may need to be performed every six months; if used infrequently an 
annual check may suffice. 

6.1.1. 

Leak Test 

Attach  the  flowmeter  to  a  low  pressure  medical  gas  supply  with  a  nominal  supply  pressure  as 
specified on the rating label of the device and close the flow control valve.  Connect supply tubing 
to the outlet barb and immerse the other end of the tube in water, the presence of gas bubbles in 
the water indicates a leak.  Remove the tube and apply an oxygen compatible leak test solution to 
all outlets and fittings and check for leaks.  A flowmeter that fails these tests should be removed 
from use. 

6.1.2. 

Flow Test 

Verify flow rates at all flow settings against those given in the Device Specification in section 6. 
Mass flow meters with appropriate full-scale ranges for the flow rates are particularly suitable for 
this. 

Alternatively,  where  flow  measuring  equipment  is  not  available,  a  qualitative  check  can  be 
performed by confirming that gas flow can be sensed to be increasing with each step increase in 
flow across the full flow range. For higher flow rates, this can be sensed audibly or by holding a 
hand close to the flow outlet; moistening the skin can increase the cooling sensation and therefore 
sensitivity to the flow. 

For very flow rates, this may better be achieved by placing the end of a connected tube in a glass 
of water to observe the bubbles. 

Summary of Contents for 829-0301

Page 1: ...702 0049 22 September 2020 Firesafe Flowmeter Instructions for Use...

Page 2: ......

Page 3: ...6 4 1 Connecting to Gas Supply 6 4 2 Adjusting the Flow 6 4 3 Reading the Flowmeter 6 4 4 Connecting to Patient 6 4 5 During Use 7 4 6 After Use 7 5 Cleaning and Disinfection 7 5 1 Cleaning 7 5 2 Sus...

Page 4: ...injury if their oxygen supply is not available or is stopped unexpectedly The flowmeter is not MRI compatible Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied...

Page 5: ...sition Failure to do so may result in incorrect flow rates being displayed Arrange the oxygen tubing and gas hose carefully to avoid damage to the hose and the potential for causing a trip hazard Neve...

Page 6: ...vices 3 2 Technical Description The Firesafe enables accurate control of the flow of either medical oxygen or medical air model dependent in the range of 1 to 15 l min A precision rotating float rises...

Page 7: ...fluted rim which continuously rotates in the gas stream to provide visual confirmation that gas is flowing and the device is providing an accurate reading Historically incidents have occurred with co...

Page 8: ...connection before turning on the supply pressure Do not use the device to tighten or lock the connection For quick connector probes e g BS 5682 DIN AFNOR ensure that the connection is correctly made b...

Page 9: ...formed every six months if used infrequently an annual check may suffice 6 1 1 Leak Test Attach the flowmeter to a low pressure medical gas supply with a nominal supply pressure as specified on the ra...

Page 10: ...n flow rates different from the flow settings Supply Pressure Range Minimum 370 kPa Maximum 550 kPa Accuracy Nominal Pressure 10 of indicated reading between 1 15 l min Valve Settling 2 from initial s...

Page 11: ...to operate the medical gas pipeline system isolation valves Inlet filter blocked Repair required Flowsafe Valve has activated Flowsafe Valve is designed to activate when an exceptionally high flow of...

Page 12: ...Flowmeter O2 Single Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0902 Firesafe Flowmeter O2 Dual Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0951 Firesafe Flowmeter Air Single Direct 0...

Page 13: ...er Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1...

Page 14: ...12...

Page 15: ...13...

Page 16: ...BPR Medical Limited 44 0 1623 628 281 Qarad EC REP BV 22 Hamilton Way Mansfield Notts info bprmedical com Pas 257 2440 Geel NG18 5BU United Kingdom www bprmedical com Belgium...

Reviews: