Firesafe 829-0301 Instructions For Use Manual Download Page 4

 

 1.

 

Symbols 

Warning! 

Indicates a potentially hazardous situation which, if not avoided, could result in 
personal injury to the user or others 

Caution! 

Indicates a potentially hazardous situation which, if not avoided, could result in 
damage to the equipment or property 

Note 

Highlights points that might allow more convenient or efficient operation of the 
equipment 

 

Use no oil 

 

Flow should be read level with the top of the float 

 2.

 

Warnings and Cautions 

2.1. 

Warnings! 

 

Read  through  this  entire  instruction  manual  before  using  or  showing  others  how  to  use  this 
flowmeter.    As  with  all  medical  equipment,  attempting  to  use  this  device  without  a  thorough 
understanding of its operation may result in patient or user injury. 

 

Medical gases are, or should be considered a drug and should only be used for medical purposes 
as prescribed by a physician or authorised clinician and in accordance with their instructions. 

 

Oxygen  therapy  may  be  a  critical  treatment.    Additional  risk  control  measures  such  as  SpO2 
monitoring with an alarm and a backup oxygen supply with delivery device should be provided for 
patients  who  might  suffer  distress  or  injury  if  their  oxygen  supply  is  not  available  or  is  stopped 
unexpectedly. 

 

The flowmeter is not MRI compatible. 

 

Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied within 
the pressure range given in the Device Specification.  If the supply is a gas cylinder, check the 
cylinder contents gauge regularly. 

 

This  flowmeter  is  only  for  use  with  the  medical  gas  indicated  on  the  device  label.    Check  the 
cylinder or supply is the correct medical grade gas before use.  Always check that the flowmeter 
is connected to the correct gas specific terminal unit. 

 

Gas specific connectors are fitted to the flowmeter.  Do not attempt to modify the fittings to suit 
other gases or fitting systems. 

 

Oxygen is not flammable; however the presence of it will drastically increase the rate and severity 
of combustion.  Oil and/or grease in the presence of an oxygen enriched atmosphere will become 
highly combustible.  Oxygen must never be allowed to come into contact with oil, grease or other 
petroleum-based substances.  Do not use oil or grease on this flowmeter. 

 

Many  hand  creams  and  moisturisers  contain  paraffin  and  petroleum  bases  which  are  highly 
flammable and must never be allowed to contact the Firesafe Flowmeter. Ensure hands are clean 
and dry before operating the equipment. 

 

Do not smoke around oxygen equipment. 

Summary of Contents for 829-0301

Page 1: ...702 0049 22 September 2020 Firesafe Flowmeter Instructions for Use...

Page 2: ......

Page 3: ...6 4 1 Connecting to Gas Supply 6 4 2 Adjusting the Flow 6 4 3 Reading the Flowmeter 6 4 4 Connecting to Patient 6 4 5 During Use 7 4 6 After Use 7 5 Cleaning and Disinfection 7 5 1 Cleaning 7 5 2 Sus...

Page 4: ...injury if their oxygen supply is not available or is stopped unexpectedly The flowmeter is not MRI compatible Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied...

Page 5: ...sition Failure to do so may result in incorrect flow rates being displayed Arrange the oxygen tubing and gas hose carefully to avoid damage to the hose and the potential for causing a trip hazard Neve...

Page 6: ...vices 3 2 Technical Description The Firesafe enables accurate control of the flow of either medical oxygen or medical air model dependent in the range of 1 to 15 l min A precision rotating float rises...

Page 7: ...fluted rim which continuously rotates in the gas stream to provide visual confirmation that gas is flowing and the device is providing an accurate reading Historically incidents have occurred with co...

Page 8: ...connection before turning on the supply pressure Do not use the device to tighten or lock the connection For quick connector probes e g BS 5682 DIN AFNOR ensure that the connection is correctly made b...

Page 9: ...formed every six months if used infrequently an annual check may suffice 6 1 1 Leak Test Attach the flowmeter to a low pressure medical gas supply with a nominal supply pressure as specified on the ra...

Page 10: ...n flow rates different from the flow settings Supply Pressure Range Minimum 370 kPa Maximum 550 kPa Accuracy Nominal Pressure 10 of indicated reading between 1 15 l min Valve Settling 2 from initial s...

Page 11: ...to operate the medical gas pipeline system isolation valves Inlet filter blocked Repair required Flowsafe Valve has activated Flowsafe Valve is designed to activate when an exceptionally high flow of...

Page 12: ...Flowmeter O2 Single Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0902 Firesafe Flowmeter O2 Dual Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0951 Firesafe Flowmeter Air Single Direct 0...

Page 13: ...er Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1...

Page 14: ...12...

Page 15: ...13...

Page 16: ...BPR Medical Limited 44 0 1623 628 281 Qarad EC REP BV 22 Hamilton Way Mansfield Notts info bprmedical com Pas 257 2440 Geel NG18 5BU United Kingdom www bprmedical com Belgium...

Reviews: