Firesafe 829-0301 Instructions For Use Manual Download Page 10

 

6.2. 

Service and Repair 

Direct connection flowmeters do not require routine servicing but can be repaired.  The device can 
be  returned  to  a  recognised  BPR  Medical  Service  Centre,  details  of  which  can  be  found  at 

www.bprmedical.com

.

 

Indirect  connection  flowmeters  should  have  the  hose  assembly  replaced  after  four  years  use  or 
five years from the date of manufacture.  The manufacture date can be found on the device label.  
Device life can be calculated by adding 7 years to the date of manufacture. 

Disassembly, repair or reassembly must be carried out by a suitably qualified person working in a 
controlled environment. 

 7.

 

Specification 

Specification 

Value 

Flow Scale 

0 – 15 l/min 

Inlet Pressure Range 

Model specific between 370 – 550 kPa 

Note: The flowmeter is able to operate safely within the pressure range 370 kPa 
to 550 kPa, but the flowmeter will only deliver accurately when connected to an 
input pressure as shown on the device.  Any input pressure other than shown on 
the device will result in flow rates different from the flow settings 

Supply Pressure Range

 

Minimum 370 kPa 

Maximum 550 kPa 

Accuracy 

Nominal Pressure: ±10 % of indicated reading between 1 – 15 l/min 

Valve Settling: < 2 % from initial setting within 3 hours 

Temperature: 3 % flow increase/decrease per 5 °C temperature increase/decrease

Inlet Pressure: 4 % increase/decrease per 10 kPa increase/decrease from nominal 
pressure 

Outlet Flow Resistance: Negligible up to 230 kPa gauge back pressure 

Gas Compatibility 

Medical Oxygen or Medical Air (model dependant) 

Maximum Flow (Flush Flow 
Mode) 

25 – 30 l/min 

Firesafe™ Nozzle Flow Range  1 – 25 l/min 

Inlet Filtration 

25 µm 

Environmental 

Transport and Storage Temperature: -20 °C to 60 °C (-4 °F to 140 °F) 

Operating Temperature: 0 °C to 40 °C (32 °F to 104 °F) 

Humidity: 0-100% RH non-condensing 

Regulatory 

CE: Medical Device Directive 93/42/EEC – Active Medical Device – Class IIa 

Applied Standards 

BS EN ISO 5359 

Low pressure hose assemblies for use with medical gases 

BS EN ISO 15001 

Anaesthetic and respiratory equipment.  Compatibility with oxygen 

BS EN ISO 15002 

Flow-metering devices for connection to terminal units of medical gas pipeline 
systems 

BS EN ISO 19054 

Rail systems for supporting medical equipment 

Summary of Contents for 829-0301

Page 1: ...702 0049 22 September 2020 Firesafe Flowmeter Instructions for Use...

Page 2: ......

Page 3: ...6 4 1 Connecting to Gas Supply 6 4 2 Adjusting the Flow 6 4 3 Reading the Flowmeter 6 4 4 Connecting to Patient 6 4 5 During Use 7 4 6 After Use 7 5 Cleaning and Disinfection 7 5 1 Cleaning 7 5 2 Sus...

Page 4: ...injury if their oxygen supply is not available or is stopped unexpectedly The flowmeter is not MRI compatible Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied...

Page 5: ...sition Failure to do so may result in incorrect flow rates being displayed Arrange the oxygen tubing and gas hose carefully to avoid damage to the hose and the potential for causing a trip hazard Neve...

Page 6: ...vices 3 2 Technical Description The Firesafe enables accurate control of the flow of either medical oxygen or medical air model dependent in the range of 1 to 15 l min A precision rotating float rises...

Page 7: ...fluted rim which continuously rotates in the gas stream to provide visual confirmation that gas is flowing and the device is providing an accurate reading Historically incidents have occurred with co...

Page 8: ...connection before turning on the supply pressure Do not use the device to tighten or lock the connection For quick connector probes e g BS 5682 DIN AFNOR ensure that the connection is correctly made b...

Page 9: ...formed every six months if used infrequently an annual check may suffice 6 1 1 Leak Test Attach the flowmeter to a low pressure medical gas supply with a nominal supply pressure as specified on the ra...

Page 10: ...n flow rates different from the flow settings Supply Pressure Range Minimum 370 kPa Maximum 550 kPa Accuracy Nominal Pressure 10 of indicated reading between 1 15 l min Valve Settling 2 from initial s...

Page 11: ...to operate the medical gas pipeline system isolation valves Inlet filter blocked Repair required Flowsafe Valve has activated Flowsafe Valve is designed to activate when an exceptionally high flow of...

Page 12: ...Flowmeter O2 Single Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0902 Firesafe Flowmeter O2 Dual Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0951 Firesafe Flowmeter Air Single Direct 0...

Page 13: ...er Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1...

Page 14: ...12...

Page 15: ...13...

Page 16: ...BPR Medical Limited 44 0 1623 628 281 Qarad EC REP BV 22 Hamilton Way Mansfield Notts info bprmedical com Pas 257 2440 Geel NG18 5BU United Kingdom www bprmedical com Belgium...

Reviews: