Operating Instructions 

4.1. 

Connecting to Gas Supply 

Before  use,  visually  check  both  the  flowmeter  and  hose  (if  fitted)  for  any  damage  or 
contamination.  Do not connect or use the device if there are any doubts about its condition. 

The flowmeter is supplied with a gas specific connector that is designed to connect to a mating 
gas specific outlet.  Gas specific outlets might be a terminal unit in a medical gas pipeline system 
or part of a pressure regulator outlet on a gas cylinder. 

If  you  are  using  a  cylinder  supply,  ensure  that  the  cylinder  contents  are  adequate  for  planned 
therapy and turn on the supply at the cylinder. 

Connect the gas specific quick connect to the appropriate gas specific outlet. 

For indirect connection flowmeters the Medirail bracket should be fastened to a suitable medical 
rail  system  (compliant  with  BS  EN  ISO  19054  Rail  systems  for  supporting  medical  equipment) 
using the thumb screw in the Medirail bracket. 

Warning!

 Where the gas specific connector is dependent on a threaded fastener (e.g. DISS CGA 

– V5 1240, AS 2896/SIS handwheel) offer device to outlet and connect a few turns.  Align device 
to  the  final  vertical  position  and  fully  hand  tighten  the  connection  before  turning  on  the  supply 
pressure.  Do not use the device to tighten or lock the connection. 

For quick connector probes (e.g. BS 5682, DIN, AFNOR), ensure that the connection is correctly 
made  by  gently  pulling  the  flowmeter  body  or  hose  (as  applicable) before  turning  on  the  supply 
pressure. 

Connect a sufficient length of patient tubing (not supplied) to the outlet barb. 

Note: 

Removal  of  the  Firesafe™  Nozzle  when  fitting  a  humidifier  will  temporarily  disable  the 

oxygen arresting capability of the flowmeter in the event of a cannula fire.  Always remember to 
refit  the  Firesafe™  Nozzle  (indicated  by  a  distinctive  orange  band)  when  the  humidifier  is  no 
longer needed. 

4.2. 

Adjusting the Flow 

Turn the flow control valve anti-clockwise to turn on and increase the flow of gas. 

If the flowmeter does not operate correctly, remove it from use and refer to the trouble shooting 
guide in Section 8 of this booklet. 

4.3. 

Reading the Flowmeter 

The  diagram  opposite  shows  the  correct  reading 
level  in  relation  to  the  float  position  for  both  float 
type and ball type variable area flowmeters. 

The  right  hand  illustration  indicates  the  correct 
reading  level  for  this  Firesafe™  Flowmeter  i.e.  at 
the top of the float. 

4.4. 

Connecting to Patient 

Connect  the  other  end  of  the  oxygen  tubing  to  the  patient  or  patient  enclosure  using  the 
appropriate connector (not supplied). 

Reading Level

Float

Ball

Summary of Contents for 829-0301

Page 1: ...702 0049 22 September 2020 Firesafe Flowmeter Instructions for Use...

Page 2: ......

Page 3: ...6 4 1 Connecting to Gas Supply 6 4 2 Adjusting the Flow 6 4 3 Reading the Flowmeter 6 4 4 Connecting to Patient 6 4 5 During Use 7 4 6 After Use 7 5 Cleaning and Disinfection 7 5 1 Cleaning 7 5 2 Sus...

Page 4: ...injury if their oxygen supply is not available or is stopped unexpectedly The flowmeter is not MRI compatible Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied...

Page 5: ...sition Failure to do so may result in incorrect flow rates being displayed Arrange the oxygen tubing and gas hose carefully to avoid damage to the hose and the potential for causing a trip hazard Neve...

Page 6: ...vices 3 2 Technical Description The Firesafe enables accurate control of the flow of either medical oxygen or medical air model dependent in the range of 1 to 15 l min A precision rotating float rises...

Page 7: ...fluted rim which continuously rotates in the gas stream to provide visual confirmation that gas is flowing and the device is providing an accurate reading Historically incidents have occurred with co...

Page 8: ...connection before turning on the supply pressure Do not use the device to tighten or lock the connection For quick connector probes e g BS 5682 DIN AFNOR ensure that the connection is correctly made b...

Page 9: ...formed every six months if used infrequently an annual check may suffice 6 1 1 Leak Test Attach the flowmeter to a low pressure medical gas supply with a nominal supply pressure as specified on the ra...

Page 10: ...n flow rates different from the flow settings Supply Pressure Range Minimum 370 kPa Maximum 550 kPa Accuracy Nominal Pressure 10 of indicated reading between 1 15 l min Valve Settling 2 from initial s...

Page 11: ...to operate the medical gas pipeline system isolation valves Inlet filter blocked Repair required Flowsafe Valve has activated Flowsafe Valve is designed to activate when an exceptionally high flow of...

Page 12: ...Flowmeter O2 Single Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0902 Firesafe Flowmeter O2 Dual Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0951 Firesafe Flowmeter Air Single Direct 0...

Page 13: ...er Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1...

Page 14: ...12...

Page 15: ...13...

Page 16: ...BPR Medical Limited 44 0 1623 628 281 Qarad EC REP BV 22 Hamilton Way Mansfield Notts info bprmedical com Pas 257 2440 Geel NG18 5BU United Kingdom www bprmedical com Belgium...

Reviews: