3.2.1. 

Firesafe™ Nozzle 

Every Firesafe™ Flowmeter is fitted with a Firesafe™ Nozzle, minimising the known risks of fire 
associated  with  oxygen  therapy.    The  Firesafe™  Nozzle  is  discernible  from  a  standard  tubing 
connector as it features a bright orange band.  The Firesafe™ Nozzle can limit the impact of an 
oxygen fire by automatically arresting the oxygen flow.  Acting like a fuse, it automatically isolates 
the oxygen supply as a fire tracks back along the tubing towards the oxygen source.  Activation of 
the Firesafe™ Nozzle during a fire requires is highly reliable at oxygen flows above 1 l/min.  The 
Firesafe™ Nozzle is not suitable for use as a fire risk control measure outside this range. 

3.2.2. 

Flowsafe™ Valve 

The Flowsafe™ Valve is another feature unique to the Firesafe™ Flowmeter that further lowers 
the  risk  of  fire  associated  with  oxygen  use.    In  the  event  of  the  flow  tube  being  inadvertently 
sheared from the flowmeter body, it automatically isolates the oxygen flow, preventing a free flow 
of  oxygen  into  the  hospital  ward.    The  pipeline  pressure  is  therefore  maintained,  negating  the 
need to isolate the ward oxygen supply, upon which other patients may be dependent. 

3.2.3. 

Precision Rotating Float 

Unlike  conventional  ball  and  glass  style  flowmeters,  the  Firesafe™  Flowmeter  incorporates  a 
precision  float  with  a  fluted  rim,  which  continuously  rotates  in  the  gas  stream  to  provide  visual 
confirmation  that  gas  is  flowing  and  the  device  is  providing  an  accurate  reading.    Historically, 
incidents have occurred with conventional designs whereby static charge within the flow tube has 
caused the ball to stick to the surface and provide a false reading. 

Unlike conventional ball style flowmeters, where there is often doubt as to where a reading should 
be  taken,  the  Precision  Rotating  Float  has  a  red  band  indicating  the  position  at  which  the  flow 
reading should be taken. 

3.2.4. 

Static Control (Triboelectric Charge Dissipation) 

The  Firesafe™  Flowmeter  incorporates  two  separate  static  control  measures  in  order  to  rapidly 
dissipate  any  charge  generated  during  use.    These  unique  measures  control  the  recognised 
phenomena of static charge, which is known to cause the ball in conventional flowmeters to stick 
to the flow tube inner surface, leading to either an inaccurate reading or indicating a flow exists 
when the valve is closed. 

3.2.5. 

Flush Flow Mode 

The  flowmeter  has  a  Flush  Flow  Mode.  This  mode  may  be  used  to  deliver  approximately  
25-30 l/min to supply a resuscitation bag. 

3.2.6. 

Inlet Filter 

The  flowmeter  is  fitted  with  a  25  μm  filter  to  ensure  that  large  particles  of  debris  that  might  be 
resident in the gas supply do not contaminate the flowmeter or pass through to the patient. 

3.2.7. 

Gas Specific Connector 

The  gas  specific  quick  connection  probe  has  characteristics  to  prevent  connections  between 
different gas services. 

 

 

Summary of Contents for 829-0301

Page 1: ...702 0049 22 September 2020 Firesafe Flowmeter Instructions for Use...

Page 2: ......

Page 3: ...6 4 1 Connecting to Gas Supply 6 4 2 Adjusting the Flow 6 4 3 Reading the Flowmeter 6 4 4 Connecting to Patient 6 4 5 During Use 7 4 6 After Use 7 5 Cleaning and Disinfection 7 5 1 Cleaning 7 5 2 Sus...

Page 4: ...injury if their oxygen supply is not available or is stopped unexpectedly The flowmeter is not MRI compatible Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied...

Page 5: ...sition Failure to do so may result in incorrect flow rates being displayed Arrange the oxygen tubing and gas hose carefully to avoid damage to the hose and the potential for causing a trip hazard Neve...

Page 6: ...vices 3 2 Technical Description The Firesafe enables accurate control of the flow of either medical oxygen or medical air model dependent in the range of 1 to 15 l min A precision rotating float rises...

Page 7: ...fluted rim which continuously rotates in the gas stream to provide visual confirmation that gas is flowing and the device is providing an accurate reading Historically incidents have occurred with co...

Page 8: ...connection before turning on the supply pressure Do not use the device to tighten or lock the connection For quick connector probes e g BS 5682 DIN AFNOR ensure that the connection is correctly made b...

Page 9: ...formed every six months if used infrequently an annual check may suffice 6 1 1 Leak Test Attach the flowmeter to a low pressure medical gas supply with a nominal supply pressure as specified on the ra...

Page 10: ...n flow rates different from the flow settings Supply Pressure Range Minimum 370 kPa Maximum 550 kPa Accuracy Nominal Pressure 10 of indicated reading between 1 15 l min Valve Settling 2 from initial s...

Page 11: ...to operate the medical gas pipeline system isolation valves Inlet filter blocked Repair required Flowsafe Valve has activated Flowsafe Valve is designed to activate when an exceptionally high flow of...

Page 12: ...Flowmeter O2 Single Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0902 Firesafe Flowmeter O2 Dual Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0951 Firesafe Flowmeter Air Single Direct 0...

Page 13: ...er Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1...

Page 14: ...12...

Page 15: ...13...

Page 16: ...BPR Medical Limited 44 0 1623 628 281 Qarad EC REP BV 22 Hamilton Way Mansfield Notts info bprmedical com Pas 257 2440 Geel NG18 5BU United Kingdom www bprmedical com Belgium...

Reviews: