Only  appropriately  trained  personnel  working  in  controlled  conditions  may  disassemble  or 
assemble this flowmeter. 

 

Do not disassemble the flowmeter whilst under pressure. 

 

Secure cylinders to a wall, stand or cart in accordance with local safety regulations. 

 

Do not stand in front of a flowmeter or regulator outlet when opening a cylinder valve. 

 

Do not allow any fluid to enter this device. 

 

The  accuracy  of  the  flowmeter  can  be  affected  significantly  if  the  input  pressure  is  other  than 
stated in the technical specification and the device label. 

 

Do not directly connect the flowmeter output to another pressurised gas source as this may lead 
to corruption of one or other of the gas supplies (e.g. connecting an oxygen flowmeter to an air 
flowmeter via a Y-piece in an attempt to blend their outputs). 

 

Only  use  this  flowmeter  with  the  flow  tube  in  a  vertical  position.    Failure  to  do  so may  result  in 
incorrect flow rates being displayed. 

 

Arrange the oxygen tubing and gas hose carefully to avoid damage to the hose and the potential 
for causing a trip hazard.  Never pull or apply excessive force to the gas hose.  A leaking hose 
may result in high local oxygen and nitrous oxide concentrations and an increased risk of fire. 

 

Gas flow may be prevented if the supply tubing is kinked or occluded. 

 

Care should be taken to ensure that the flow control valve is free from obstruction and unintended 
pressure.  These scenarios may result in temporary changes in flow. 

2.2. 

Cautions! 

 

The performance of the flowmeter may be affected if it is stored or transported in temperatures 
outside of the range -20 °C to 60 °C (-4 °F to 140 °F). 

 

The flowmeter is not suitable for autoclaving. 

 

Where a threaded fastener is used, do not use the device to tighten or lock connection. 

 

The performance of the flowmeter may be affected if the flow control valve is over-tightened when 
switching off flow.  Apply the minimum force required to switch off gas flow. 

 

 

Summary of Contents for 829-0301

Page 1: ...702 0049 22 September 2020 Firesafe Flowmeter Instructions for Use...

Page 2: ......

Page 3: ...6 4 1 Connecting to Gas Supply 6 4 2 Adjusting the Flow 6 4 3 Reading the Flowmeter 6 4 4 Connecting to Patient 6 4 5 During Use 7 4 6 After Use 7 5 Cleaning and Disinfection 7 5 1 Cleaning 7 5 2 Sus...

Page 4: ...injury if their oxygen supply is not available or is stopped unexpectedly The flowmeter is not MRI compatible Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied...

Page 5: ...sition Failure to do so may result in incorrect flow rates being displayed Arrange the oxygen tubing and gas hose carefully to avoid damage to the hose and the potential for causing a trip hazard Neve...

Page 6: ...vices 3 2 Technical Description The Firesafe enables accurate control of the flow of either medical oxygen or medical air model dependent in the range of 1 to 15 l min A precision rotating float rises...

Page 7: ...fluted rim which continuously rotates in the gas stream to provide visual confirmation that gas is flowing and the device is providing an accurate reading Historically incidents have occurred with co...

Page 8: ...connection before turning on the supply pressure Do not use the device to tighten or lock the connection For quick connector probes e g BS 5682 DIN AFNOR ensure that the connection is correctly made b...

Page 9: ...formed every six months if used infrequently an annual check may suffice 6 1 1 Leak Test Attach the flowmeter to a low pressure medical gas supply with a nominal supply pressure as specified on the ra...

Page 10: ...n flow rates different from the flow settings Supply Pressure Range Minimum 370 kPa Maximum 550 kPa Accuracy Nominal Pressure 10 of indicated reading between 1 15 l min Valve Settling 2 from initial s...

Page 11: ...to operate the medical gas pipeline system isolation valves Inlet filter blocked Repair required Flowsafe Valve has activated Flowsafe Valve is designed to activate when an exceptionally high flow of...

Page 12: ...Flowmeter O2 Single Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0902 Firesafe Flowmeter O2 Dual Direct 0 15 l min SIS connector AS 2896 400 kPa 829 0951 Firesafe Flowmeter Air Single Direct 0...

Page 13: ...er Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1...

Page 14: ...12...

Page 15: ...13...

Page 16: ...BPR Medical Limited 44 0 1623 628 281 Qarad EC REP BV 22 Hamilton Way Mansfield Notts info bprmedical com Pas 257 2440 Geel NG18 5BU United Kingdom www bprmedical com Belgium...

Reviews: