4. The measured body temperature is higher than 42.9°
C or lower than 32.0 ° C
。
5. The measured object is not correct.
The screen shows "Er1”
1. The environment temperature is exceed the normal
working arrange which is 15° C ~ 35 ° C (59° F to 95 ° F)
Buttons have no response
2. Load and unload the battery
3. Check if the setting is in the progress
No display or display
abnormal
Unload the battery and load again
Power off when open
Check the battery, load and unload the battery again
13. EMC information
Table 5
Guidance and manufacturer ’s declaration-electromagnetic emissions
The FC-IR101 is intended for using in the electromagnetic environment specified below. The customer or the user
of the FC-IR101 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The FC-IR101 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause and
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The FC-IR101 is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations /flicker
emissions
IEC 61000-3-3
N/A
Table 6
Guidance and manufacturer ’s declaration-electromagnetic immunity
The FC-IR101 is intended for use in the electromagnetic environment specified below. The customer or the user of
the FC-IR101 should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Complia
nce level
Electromagnetic
environment -guidance