Braun ThermoScan® PRO 6000 

Thermometer Inservice 

Summary of Contents for ThermoScan PRO 6000

Page 1: ...Braun ThermoScan PRO 6000 Thermometer Inservice ...

Page 2: ...essories 4 Approved Cleaning Solutions 5 Probe Covers 6 Taking a Temperature 7 Interferences 8 Contraindications 9 Memory and Temperature Scale 10 Manual Timer 11 Error Messages and Solutions 12 Batteries 13 Cradles and Charging Station 14 Security 15 Security Function 16 Resources ...

Page 3: ... ideal by correcting for shallow placement ExacTemp light to ensure proper steady probe positioning Pre heated Probe Tip for more accurate and faster readings Displayed Temperature Range 68 108 F 20 42 2 C Age Range Full term newborns pediatrics and adults Secure Customizable security features to prevent theft including security tether and wall mountable cradle ...

Page 4: ...yl alcohol only Use gentle pressure With probe facing down wipe the probe with a damp cloth or cleaning wipe moistened with isopropyl or ethyl alcohol Gently wipe with dry clean cotton swab or cloth immediately Allow at least 5 minutes drying time before taking temperature Make sure probe lens window is clean and dry prior to use ...

Page 5: ...probe to protect this area when cleaning the body of the thermometer Use a damp cloth or cleaning wipe with cleaning solution from the Approved Cleaning Solutions Table To clean the body ensure that the wipe is moist not saturated Wipe with display facing up Clean accessories using same method Dry 5 min ...

Page 6: ...Approved Cleaning Solutions ...

Page 7: ...nto the cradle Probe covers Are intended for single use only Always use PRO 6000 disposable probe covers ONLY Damaged perforated or soiled covers can lead to inaccurate temperatures The device will not take a temperature without a probe cover in place ...

Page 8: ...Taking a Temperature Proper Probe Placement Guide ...

Page 9: ...Taking a Temperature NOTE If no action is taken the thermometer will enter SLEEP after 10 seconds or upon returning to the cradle or host device ...

Page 10: ...Interferences For persons wearing hearing aids or in extreme ambient temperatures remove the hearing aid or the individual from the situation and wait 20 minutes prior to taking a temperature ...

Page 11: ...canal Not intended for pre term babies or small for gestational age babies Complete ear canal occlusion due to cerumen ear wax can result in lower temperature readings Use the untreated ear if prescription ear drops or other ear medications have been placed in the ear canal Patients who have deformities of the face and or ear may not be able to have a temperature taken with an ear thermometer ...

Page 12: ...tton is pushed again a new probe cover is applied or the thermometer enters sleep state Memory can also be accessed from the sleep mode and will display for 5 seconds Temperature Scale Switch to alternative temperature scale by pushing the C F button while a reading is being displayed Press C F button again to revert back Also reverts back when unit goes to sleep ...

Page 13: ...second to activate timer a beep is heard at initiation of timer 2 The timer will beep at the completion of each 15 second interval Then the current segment will become solid and the next segment will flash 3 At 60 seconds a long beep will sound all quadrants will appear solid ending the timer function 4 The thermometer will exit the timer mode after an additional 5 seconds 5 To stop the timer at a...

Page 14: ...Error Messages and Solutions ...

Page 15: ...Error Messages and Solutions ...

Page 16: ...t Can use rechargeable batteries 700 measurements per charge Requires Charging Station sold separately If needed change the C F default setting with switch in the battery compartment using a pen or pointed object To install the AA alkaline batteries Rechargeable battery pack ...

Page 17: ...Wall mountable holds 2 boxes probe covers Optional Charging Station available separately For use with rechargeable batteries Fully recharges in approximately 3 4 hours from empty Fast Charge Nickel metal hydride battery type Up to 700 measurements from a full charge Also holds a sleeve of 200 probe covers Two Protective Cradles available to choose from Small Cradle Large Cradle ...

Page 18: ...t security tethers screw to the cradle and thermometer Tether also attaches to Connex Spot Monitor Electronic Time Security see next slide Select 1 12 hours operation until thermometer locks To unlock dock the device and cradle into any Charging Station Requires Charging Station or Connex Spot Monitor and Service Tool software to activate security function ...

Page 19: ...ning next to the unlocked icon At 5 minutes prior to expiration the thermometer will begin to beep once per minute and the unlocked icon will flash 3 Upon expiration of the time limit the unlocked icon and countdown timer will be replaced with the locked icon and the thermometer will beep once every 5 minutes to indicate locked state Unlock the thermometer by returning it to the charging station N...