background image

 

 

 

 

 

 

 

 

                                           

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for PT2L

Page 1: ......

Page 2: ...ON 3 LCD SCREEN INSTRUCTIONS 3 PRODUCT PERFORMANCE 4 INTENDED USE 4 CONTRAINDICATIONS 5 INSTRUCTION FOR USE 5 PRODUCT ERRORS AND TROUBLESHOOTING 8 BODY TEMPERATURE 8 CARE AND CLEANING 9 MAINTENANCE 9 INCLUDED IN DELIVERY 10 SIGNS AND SYMBOLS 10 OTHER STANDARDS AND COMPLIANCES 10 ELECTROMAGNETIC COMPATIBILITY INFORMATION 11 WARRANTY 12 ...

Page 3: ...fore measuring 7 Patients should not drink eat or be physically active before while taking a reading Wait 30 minutes before taking a measurement Temperature readings taken when a body is in a state of stable equilibrium is more accurate and useful as a reference 8 Do not take temperature measurement over scar tissue open sores or abrasions 9 If there is a temperature difference between the thermom...

Page 4: ...ctions are unknown 18 The patient is an intended operator 19 This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installe...

Page 5: ...3 Overall Description The thermometer is mainly comprised of a plastic housing IR temperature sensor LCD display screen and batteries LCD screen instructions ...

Page 6: ... Humidity 95 RH non condensing Atmospheric Pressure 70KPa 106KPa 10 Transportation storage conditions Temperature 25 55 13 131 Humidity 95 RH non condensing Atmospheric Pressure 70KPa 106KPa 11 Display LCD display 12 Operation mode adjusted mode reference body site oral 13 Expected service life 5 years 14 Interval between 2 measurements at least 5 seconds 15 Battery life approx 3000 times measurem...

Page 7: ...d power off automatically if the LCD displays then replace the batteries to ensure ample power supply Note The batteries and electronic instruments must be disposed of in accordance with the locally applicable regulation not with domestic waste 2 Measurement process 1 Aim the thermometer probe at the center of the forehead and keep the probe no more than 3cm 1 18in away from the forehead the optim...

Page 8: ...ing is required simply let the device sit idle for 8 seconds to power off automatically 4 Buzzer Prompt Switch 1 Under the power off state press and hold the measurement button until the buzzer symbol appears alternately on the screen 2 Release the button when appears buzzer function open 3 Release the button when disappears buzzer function close 4 After setting the buzzer prompt switch the screen...

Page 9: ...tch OFF after 8 seconds To continue using the device old batteries must be replaced 8 Replacing batteries 1 Press down and hold the battery cover with the finger and apply some force to slide the cover backwards to open the battery compartment 2 Remove the old batteries and install the new batteries 3 Refer to the battery polarity symbols to orient the batteries properly during installation Make s...

Page 10: ...tate All symbols are flashing on the screen The product is not usable Please contact customer services Body temperature Body temperature runs approximately from 35 5 37 8 95 9 F to100 F To determine if one has a fever compare the temperature detected with the person s normal temperature A rise over the reference body temperature of 1 1 F or more is generally an indication of fever Different measur...

Page 11: ...hermometer is initially calibrated at the time of manufacture If this thermometer is used according to the use instruction periodic re adjustment is not required If any time your question the accuracy of measurement during the warranty period please contact customer services 5 No component can be maintained by user in the monitor circuit diagrams component part lists descriptions calibration instr...

Page 12: ...AL NUMBER Symbol for EUROPEAN REPRESENTATIVE IP22 The first characteristic numeral symbol for Degrees of protection against access to hazardous parts and against solid foreign objects The second characteristic numeral symbol for Degrees of protection against ingress of water Symbol for Application part type BF Symbol for COMPLIES WITH MDD93 42 EEC REQUIREMENTS OTHER STANDARDS AND COMPLIANCES This ...

Page 13: ...d in ASTM Standard E1965 98 except of clause 5 2 2 It s display range is 34 0 43 0 93 2 109 4 the Full responsibility for the conformance of this product to the standard is assumed by manufacturer ELECTROMAGNETIC COMPATIBILITY INFORMATION Table 1 Emission Phenomenon Compliance Electromagnetic environment Conducted and radiated RF emissions CISPR 11 Group 1 Class B The device is intended to be used...

Page 14: ...ar statement of defect description The warranty terms as below 1 The warranty period for device is one year from date of delivery In case of a warranty claim the date of delivery has to be proven by means of the sales receipt or invoice 2 Repairs under warranty do not extend the warranty period 3 The following cases are excluded under the warranty All damage which has arisen due to improper treatm...

Page 15: ...8 Paris France ANDON HEALTH CO LTD No 3 Jinping Street Ya An Road Nankai District Tianjin 300190 China Factory ANDON MEDICAL CO LTD Address No 26 HangYu Road Pilot Free Trade Zone Tianjin China Made in China Date of issue May 28 2020 N O 31 C hangj i ang Road N ankaiD i stri ct Ti anj i n P R C hi na EC REP A N D O N Heal th C o Ltd A UTO M A TI C BLO O D PRESSURE M O N I TO R M O D EL KD 5907 6V ...

Reviews: