Indications

The Myolab II is intended for the monitoring and display of 

electromyographic signals for the purpose of fitting the Utah Arm 

and the ProControl 2 prosthetic devices.

Contrandications

Because the Myolab II is a passive monitoring system, no known 

contraindications exist for its use.

Special Precautions

Static electricity can damage the EMG sensors of the Myolab 

II. Therefore, always pick up the preamplifier by touching the 

ground (center) electrode first. This will ensure that any static 

electrical charge present on your body is discharged to the 

ground electrode, and not to the two sensing electrodes. When 

not in use, always keep the preamplifiers inside the static 

protection bag supplied with the preamplifiers.
Never allow moisture or contaminants to enter the Myolab II or 

preamplifiers.
Avoid using the preamplifiers in an extremely moist 

environment for long periods of time. If the preamplifiers are 

installed inside the socket of a prosthesis, make sure that the 

socket provides adequate drainage for perspiration.

System Setup

Changing Batteries

Hold the Myolab II upside down with the battery compartment away 

from you. Place both thumbs on the battery door and push down 

and away from you. The compartment door will slide out. Connect 

the battery to the appropriate terminals. (Note: The use of a Duracell 

9 volt alkaline battery or equivalent is recommended.) Be sure to 

change batteries whenever the low battery indicator lights.

Connecting the Preamplifier Cable

The input preamplifier jack is located within the battery compartment 

(Figure 2). With the battery compartment door open, connect the 

preamplifier cable to the five-hole socket provided. (Note that the 

cable can be connected only one way-you cannot plug it in backwards.) 

Close the compartment door carefully, insuring that the cable 

connector is centered in the opening provided in the battery door. If 

using Triad preamplifiers, plug the Triad to Myolab Adapter (indluded) 

into the Myolab II, then plug the Triad preamplifiers into the adapter.

Summary of Contents for Myolab II

Page 1: ...Myolab II Prosthetist Manual...

Page 2: ...2...

Page 3: ...yoelectric prosthetic device senses EMG signals from intact muscles and translates these signals into usable arm movements The Myolab II uses this same advanced sensing technology to provide artifact...

Page 4: ...plifiers are installed inside the socket of a prosthesis make sure that the socket provides adequate drainage for perspiration System Setup Changing Batteries Hold the Myolab II upside down with the b...

Page 5: ...to Monitor EMG Activity 1 To avoid static damage to the preamplifiers always ensure that the patient is at the same electrical potential as the Myolab Two good ways of doing this are a After properly...

Page 6: ...he EMG signal can occur if both preamplifiers are not in contact with the patient common ground during monitoring 4 Set the gain dials Gain A and Gain B to 0 5 Rotate the sound dials A and B to the le...

Page 7: ...ired level using the appropriate sound dial Note The pitch of the sound not the volume will increase with an increase in EMG signal intensity In addition the sound of each channel differentiates the t...

Page 8: ...he same function in the sound arm e g flexing the sound elbow and contracting the biceps simultaneously in both the remnant arm and Figure 5 the sound arm Figure 5 For both bilateral and unilateral am...

Page 9: ...ic socket Identify all potential EMG control sites in this manner 6 When probing EMG control sites for the Utah Artificial Arm use the A channel of the Myolab II to monitor the muscle the A muscle mos...

Page 10: ...ck feature of the Utah Artificial Arm instruct the patient to contract both the flexor and extensor muscles simultaneously in a quick but low strength co contraction followed by relaxation the motion...

Page 11: ...control To measure the EMG signal magnitude accurately set the gain dial on 10 and the gain switch on X1 B The muscle EMG signals should be controllable without spasmodic jumping or quivering C Contra...

Page 12: ...k all connections and then turn the power On again Check the gain switch setting Change the battery If the Myolab II continues to malfunction after you have taken the above actions call the service de...

Page 13: ...voltage Bias Current 1 picoamp Filtering 1st order high pass at 10 Hz 1st order low pass at 10 kHz Output Impedance 1 Ohm Output Current 6 mA Output voltage 9 V supply voltage 7 5 V output T Cable to...

Page 14: ...14...

Page 15: ...15 Alternate leads for use with Myolab II...

Page 16: ...2018 Motion Control Inc 1930001 Rev D 10 04 2018 F Motion Control Inc 115 N Wright Brothers Drive Salt Lake City UT 84116 801 326 3434 Fax 801 978 0848 Toll Free 888 696 2767 www UtahArm com...

Reviews: