background image

9

Mark the location of the strongest EMG signal, and the total area 

in which a strong EMG is found.

4.   Instruct the patient to “contract-hold-relax” in about a three

second sequence. A smooth and even contraction is desired, 

without strenuous effort, with relaxation after each contraction. 

Instruct the patient to try to relax the muscles not being tested.
Systematically move the preamplifier by one-half inch increments 

and ask the patient to perform the “contract-hold-relax” sequence 

with the same strength of contraction for at least two repetitions 

at each location.

5.   When a potential control site is identified, mark the best

electrode location on the skin, and also mark the total area in 

which an adequate EMG signal is obtained (Figure 7). This will be 

important in locating electrodes in the prosthetic socket. Identify 

all potential EMG control sites in this manner.

6.   When probing EMG control sites for the Utah Artificial Arm, use

the “A” channel of the Myolab II to monitor the muscle (the “A” 

muscle) most appropriate to control flexion of the elbow of the 

prosthesis (usually an anterior muscle such as the biceps or 

pectoralis). This muscle is usually recommended for the hand 

closing site also, although exceptions are common, (e.g., when the 

elbow extension muscle is more easily controlled by the subject).

Figure 6

For example, relax the shoulder when contracting the biceps.
Also, the EMG signal will usually be more controllable if some 

resistance is provided to the motion of the remnant limb, either 

with the clinician’s hand or the socket of a prosthesis.

3.   Palpate the muscle as the patient contracts it and place the

preamplifier over the belly of the muscle. Touching the muscle 

helps give the subject more sensation of the muscle contraction, 

which may help in learning to control the muscle (Figure 6).

Summary of Contents for Myolab II

Page 1: ...Myolab II Prosthetist Manual...

Page 2: ...2...

Page 3: ...yoelectric prosthetic device senses EMG signals from intact muscles and translates these signals into usable arm movements The Myolab II uses this same advanced sensing technology to provide artifact...

Page 4: ...plifiers are installed inside the socket of a prosthesis make sure that the socket provides adequate drainage for perspiration System Setup Changing Batteries Hold the Myolab II upside down with the b...

Page 5: ...to Monitor EMG Activity 1 To avoid static damage to the preamplifiers always ensure that the patient is at the same electrical potential as the Myolab Two good ways of doing this are a After properly...

Page 6: ...he EMG signal can occur if both preamplifiers are not in contact with the patient common ground during monitoring 4 Set the gain dials Gain A and Gain B to 0 5 Rotate the sound dials A and B to the le...

Page 7: ...ired level using the appropriate sound dial Note The pitch of the sound not the volume will increase with an increase in EMG signal intensity In addition the sound of each channel differentiates the t...

Page 8: ...he same function in the sound arm e g flexing the sound elbow and contracting the biceps simultaneously in both the remnant arm and Figure 5 the sound arm Figure 5 For both bilateral and unilateral am...

Page 9: ...ic socket Identify all potential EMG control sites in this manner 6 When probing EMG control sites for the Utah Artificial Arm use the A channel of the Myolab II to monitor the muscle the A muscle mos...

Page 10: ...ck feature of the Utah Artificial Arm instruct the patient to contract both the flexor and extensor muscles simultaneously in a quick but low strength co contraction followed by relaxation the motion...

Page 11: ...control To measure the EMG signal magnitude accurately set the gain dial on 10 and the gain switch on X1 B The muscle EMG signals should be controllable without spasmodic jumping or quivering C Contra...

Page 12: ...k all connections and then turn the power On again Check the gain switch setting Change the battery If the Myolab II continues to malfunction after you have taken the above actions call the service de...

Page 13: ...voltage Bias Current 1 picoamp Filtering 1st order high pass at 10 Hz 1st order low pass at 10 kHz Output Impedance 1 Ohm Output Current 6 mA Output voltage 9 V supply voltage 7 5 V output T Cable to...

Page 14: ...14...

Page 15: ...15 Alternate leads for use with Myolab II...

Page 16: ...2018 Motion Control Inc 1930001 Rev D 10 04 2018 F Motion Control Inc 115 N Wright Brothers Drive Salt Lake City UT 84116 801 326 3434 Fax 801 978 0848 Toll Free 888 696 2767 www UtahArm com...

Reviews: