background image

11

Proprioceptive neuromuscular 

facilitation (PNF) exercise training 

has been found effective to develop 

remnant musculature in some 

amputees. Consult a physical or 

occupational therapist, or the Utah 

Artificial Arm Users Manual or 

Fitting Procedures Manual for more 

details.

Figure 7 - Differentiation of two muscle 

groups as visualized on Myolab II meters

10. Final Site Selection

Choose the muscle sites only after a reasonable period of training 

and practice, so that a potentially good muscle site is not rejected 

prematurely. The minimum criteria are:

A. To operate the Utah Artificial Arm, when either muscle “A” or

muscle “B” is contracted the “A” or “B” EMG signal must 

exceed the other signal (either “B” or “A”) by:

•  5 microvolts for hand control
•  12 microvolts for elbow control

To measure the EMG signal magnitude accurately, set the gain dial on 

10 and the gain switch on X1.

B. The muscle EMG signals should be controllable, without

spasmodic jumping or quivering.

C.  Contraction of the muscle should not induce pain.
D.  Scar tissue should be avoided, if possible. However, if no

other satisfactory sites are available, electrodes can be used 

over scar tissue if designed properly to avoid breakdown of 

the scar tissue.

Troubleshooting the Myolab II

1.  If you lose the EMG signal to one or both channels, or the signal is

erratic:

•  Check the connection to the Myolab II. Be sure the preamplifier 

cable (and adapter cable if used) is plugged securely into the 

five pin socket in the battery compartment.

•  Make sure the power switch is in the on position.

•  If you are measuring a very small EMG signal, the signal strength 

may be too weak for a gain switch setting of X1. Change to X10.

•  Make sure that the gain dial control is at an adequate setting.

Summary of Contents for Myolab II

Page 1: ...Myolab II Prosthetist Manual...

Page 2: ...2...

Page 3: ...yoelectric prosthetic device senses EMG signals from intact muscles and translates these signals into usable arm movements The Myolab II uses this same advanced sensing technology to provide artifact...

Page 4: ...plifiers are installed inside the socket of a prosthesis make sure that the socket provides adequate drainage for perspiration System Setup Changing Batteries Hold the Myolab II upside down with the b...

Page 5: ...to Monitor EMG Activity 1 To avoid static damage to the preamplifiers always ensure that the patient is at the same electrical potential as the Myolab Two good ways of doing this are a After properly...

Page 6: ...he EMG signal can occur if both preamplifiers are not in contact with the patient common ground during monitoring 4 Set the gain dials Gain A and Gain B to 0 5 Rotate the sound dials A and B to the le...

Page 7: ...ired level using the appropriate sound dial Note The pitch of the sound not the volume will increase with an increase in EMG signal intensity In addition the sound of each channel differentiates the t...

Page 8: ...he same function in the sound arm e g flexing the sound elbow and contracting the biceps simultaneously in both the remnant arm and Figure 5 the sound arm Figure 5 For both bilateral and unilateral am...

Page 9: ...ic socket Identify all potential EMG control sites in this manner 6 When probing EMG control sites for the Utah Artificial Arm use the A channel of the Myolab II to monitor the muscle the A muscle mos...

Page 10: ...ck feature of the Utah Artificial Arm instruct the patient to contract both the flexor and extensor muscles simultaneously in a quick but low strength co contraction followed by relaxation the motion...

Page 11: ...control To measure the EMG signal magnitude accurately set the gain dial on 10 and the gain switch on X1 B The muscle EMG signals should be controllable without spasmodic jumping or quivering C Contra...

Page 12: ...k all connections and then turn the power On again Check the gain switch setting Change the battery If the Myolab II continues to malfunction after you have taken the above actions call the service de...

Page 13: ...voltage Bias Current 1 picoamp Filtering 1st order high pass at 10 Hz 1st order low pass at 10 kHz Output Impedance 1 Ohm Output Current 6 mA Output voltage 9 V supply voltage 7 5 V output T Cable to...

Page 14: ...14...

Page 15: ...15 Alternate leads for use with Myolab II...

Page 16: ...2018 Motion Control Inc 1930001 Rev D 10 04 2018 F Motion Control Inc 115 N Wright Brothers Drive Salt Lake City UT 84116 801 326 3434 Fax 801 978 0848 Toll Free 888 696 2767 www UtahArm com...

Reviews: