background image

10

7.   Find the best agonist-antagonist pair of EMG control sites. Use

the “B” channel of the Myolab II to monitor the muscle (the 

“B” muscle) most appropriate to control powered extension 

of the elbow (and hand opening or closing). In a two-muscle 

control system such as the Utah Artificial Arm, strength and 

independence are important to the agonist-antagonist pair of 

muscle control sites.
While viewing both Myolab II meters simultaneously, probe 

for the two muscle sites with the greatest difference of EMG 

amplitude (Figure 7). When the subject contracts the “A” muscle, 

note the amplitude of co-contraction on the “B” channel. Then test 

the reverse situation as the patient contracts the “B” muscle while 

the “A” meter is monitored. Raise the gain for the weaker muscle 

using the Gain Control Dial on the Myolab II so that contraction 

of either “A” or “B” muscles results in about the same difference 

between the two meter readings.

8.   To train patients in using the unlock feature of the Utah

Artificial Arm, instruct the patient to contract both the flexor and 

extensor muscles simultaneously, in a quick but low strength 

co-contraction, followed by relaxation (the motion is similar to 

“snapping” the fingers-very quickly but without straining for 

a strong contraction). After the muscles have relaxed, the “A” 

muscle is contracted gradually to move the elbow after unlocking 

the prosthesis.

9.   Using the Myolab II to monitor activity of the appropriate

muscles, have the subject perform the following sequence 

repetitively:

Step 1- Gradually contract the “A” muscle. Maintain a 

contraction for one second and then relax.
Step 2- Gradually contract the “B” muscle. Maintain a 

contraction for one second and then relax.
Step 3- Quickly co-contract both the “A” and “B” muscles, then 

relax both. Return to Step 1 and repeat sequence.
Once the best control sites are found, the patient may use 

the Myolab II as a training device, by concentrating efforts on 

developing the maximum difference between the two Myolab 

II meters (Figure 7). Some patients may be aided by visualizing 

movements of the phantom limb while they contract the 

remnant musculature.

Summary of Contents for Myolab II

Page 1: ...Myolab II Prosthetist Manual...

Page 2: ...2...

Page 3: ...yoelectric prosthetic device senses EMG signals from intact muscles and translates these signals into usable arm movements The Myolab II uses this same advanced sensing technology to provide artifact...

Page 4: ...plifiers are installed inside the socket of a prosthesis make sure that the socket provides adequate drainage for perspiration System Setup Changing Batteries Hold the Myolab II upside down with the b...

Page 5: ...to Monitor EMG Activity 1 To avoid static damage to the preamplifiers always ensure that the patient is at the same electrical potential as the Myolab Two good ways of doing this are a After properly...

Page 6: ...he EMG signal can occur if both preamplifiers are not in contact with the patient common ground during monitoring 4 Set the gain dials Gain A and Gain B to 0 5 Rotate the sound dials A and B to the le...

Page 7: ...ired level using the appropriate sound dial Note The pitch of the sound not the volume will increase with an increase in EMG signal intensity In addition the sound of each channel differentiates the t...

Page 8: ...he same function in the sound arm e g flexing the sound elbow and contracting the biceps simultaneously in both the remnant arm and Figure 5 the sound arm Figure 5 For both bilateral and unilateral am...

Page 9: ...ic socket Identify all potential EMG control sites in this manner 6 When probing EMG control sites for the Utah Artificial Arm use the A channel of the Myolab II to monitor the muscle the A muscle mos...

Page 10: ...ck feature of the Utah Artificial Arm instruct the patient to contract both the flexor and extensor muscles simultaneously in a quick but low strength co contraction followed by relaxation the motion...

Page 11: ...control To measure the EMG signal magnitude accurately set the gain dial on 10 and the gain switch on X1 B The muscle EMG signals should be controllable without spasmodic jumping or quivering C Contra...

Page 12: ...k all connections and then turn the power On again Check the gain switch setting Change the battery If the Myolab II continues to malfunction after you have taken the above actions call the service de...

Page 13: ...voltage Bias Current 1 picoamp Filtering 1st order high pass at 10 Hz 1st order low pass at 10 kHz Output Impedance 1 Ohm Output Current 6 mA Output voltage 9 V supply voltage 7 5 V output T Cable to...

Page 14: ...14...

Page 15: ...15 Alternate leads for use with Myolab II...

Page 16: ...2018 Motion Control Inc 1930001 Rev D 10 04 2018 F Motion Control Inc 115 N Wright Brothers Drive Salt Lake City UT 84116 801 326 3434 Fax 801 978 0848 Toll Free 888 696 2767 www UtahArm com...

Reviews: