Venice evacuation chair - Venice Clack locking system
10.10 Auxiliary connection cable
The chair is supplied as standard with an auxiliary cable. This
cable is inside the bag on the rear side of the chair backrest.
It can be connected permanently (if required) to the control
unit and the motor, bypassing the wireless connection. If
permanently connected, position the cable so that it is not
crushed when opening, closing and using the chair.
The auxiliary cable allows to re-establish the wireless
connection between the control unit and the motor. The
wireless control unit and the motor are connected during
the assembly procedure and should not disconnect during
normal use.
If the connection is lost, perform the following reconnection
procedures:
1. Open the control unit connector compartment
with the aid of a coin. (Figure 10)
2. Insert the auxiliary cable connector into the port
located in the compartment (Figure 11).
3. Remove the door protection on the motor housing
(Figure 12).
4. Connect the cable's serial plug to the port on the
motor housing (Figure 13).
5. Press the power button of the chair and wait
for the wireless interface and the motor to be
synchronized. A double beep will confirm that
connection has been successfully completed.
6. Disconnect the cable from the controller by prying
it with the tip of the plug closing the connector
port (Figure 14).
7. Remove the serial cable from the motor housing
and store it in the bag provided.
Figure 10 - Opening the connector compartment
Figure 11 - Connecting the cable to the connector port
Figure 12 - Port on the motor housing
Figure 13 - Connecting the cable to the motor housing
Figure 14 - Unplugging the connector
Cable
connection port