Fema Smart DCM Operating Instructions Manual Download Page 3

SMART DCM SWITCH – OPERATING INSTRUCTIONS 

MU2B-0328GE51 R0207A 

3

PARAMETRIZATION LEVEL 

Once you have successfully entered the ENTRY CODE and 
thus gained access to the PARAMETRIZATION LEVEL, you 
can edit values and reconfigure the device. 

NOTE:

 

The currently set option (e.g. "yes," "no," "window 
monitor", etc.) is always marked in the display 
screen with a small dot to the left (see Fig. 5). 

 

Monitor

Max.

OK

Window

ESCAPE

Monitor

Max.

Window

ESCAPE

OK

DOT

 

Fig. 5. Dot indicating present setting 

 

SET-PT. 

In this screen, you can select the set-point. This is defined as 
the pressure at which you want the device to switch. 
The default setting is 60% of full-scale. 
 

ERROR MESSAGE: "VALUE OUT OF RANGE" 

If you have entered a set-point outside of the device's 
specified pressure measurement range or which is for any 
other reason not rational, this error message will appear, 
whereupon you should then enter a correct value. 
 

RESET-PT. 

In this screen, you can select the reset-point. This is defined 
as the pressure at which you want the device to switch back. 
The default setting is 40% of full-scale. 
 

ERROR MESSAGE: "VALUE OUT OF RANGE" 

If you have entered a reset-point outside of the device's 
specified pressure measurement range or which is for any 
other reason not rational, this error message will appear, 
whereupon you should then enter a correct value. 
 

OUTPUT 

MONITOR 

Here, you can choose between three types of monitors: min., 
max., and window. 

 Min.; 

 Max.; 

 Window; 

See Fig. 7 on page 5 for an explanation of the meanings of 
these different settings. 
The default setting is "Max." 
 

FUNCTION 

Here, you can choose the type of action you want the DCM 
Switch to have: 

  N.O.: Normally-open switch (meaning: when the switch is 

activated, it will CLOSE); or 

  N.C.: Normally-closed switch (meaning: when the switch 

is activated, it will OPEN). 

See Table 2 and Fig. 7 on page 5 for more information. 
The default setting is "N.O." 
 

AUTO ZERO 

The purpose of the auto zero function is to allow the user to 
adjust the device to local ambient atmospheric pressure and 

to compensate for drift. To use this function properly, the 
device must be mounted in the application and the pressure 
sensor exposed to local ambient atmospheric pressure (i.e. 
0 bar relative pressure). 

  Yes: The device takes the current local ambient 

atmospheric pressure as its zero. 

  No: The device setting remains unchanged. 

The default setting is "No". 
 

SET ZERO 

The difference between the pressure value you enter in this 
screen and the pressure which the device is currently 
measuring will be taken as the offset. This offset can be reset 
to zero using the reset zero option (see section "Reset Zero" 
below). 
The default setting is the pressure which the device is 
currently measuring (meaning that the offset is then zero). 
 

ERROR MESSAGE: "VALUE OUT OF RANGE" 

If you have set a zero outside the permissible range (±25% 
full-scale), this error message will appear, whereupon you 
should then enter a correct value. 
 

RESET ZERO 

Here, you can reset the zero: 

  Yes: The pressure offset will be set to zero. 

  No: The device setting remains unchanged. 

The default setting is "No". 
 

DROP-IN 

Here, you can select the drop-in value. See also Fig. 8. 
Possible values range from 0 to 99 seconds. 
The default setting is "0" seconds. 
 

DROP-OUT 

Here, you can select the drop-out value. See also Fig. 8.

 

Possible values range from 0 to 99 seconds. 
The default setting is "0" seconds. 
 

PEAKS 

Here, you can view the historical max. and min. measured 
pressures as well as the max. measured temperature 
(historical = measured since the device has been put into 
operation / since the last time the peaks were cleared). 
If desired, by proceeding to "Clear" and choosing "yes," you 
can also permanently erase all such values. All such peaks 
will then be erased and the recording of peaks will be re-
initiated. 
 

DISPLAY 

Here, you can select/alter various modes of displaying screen 
information. 
 

UNIT 

Here, you can choose the units in which the device displays 
pressures: 

  bar: All pressures will be displayed in bars. 

  Pa: All pressures will be displayed in Pascals. 

  Psi: All pressures will be displayed in pounds / in

2

The default setting is "bar". 
 

LANGUAGE 

English is the only supported language. 

Summary of Contents for Smart DCM

Page 1: ...0 95 r h non condensing Temperature of medium 30 100 C 16 bar models 40 100 C 16 bar models Storage temperature Versions non HMI 40 80 C 16 bar 40 100 C 16 bar Versions HMI 30 80 C Climate class Indo...

Page 2: ...here the device will check the plausibility of your values and if necessary redirect you back to the PARAMETRIZATION LEVEL and The CONFIRMATION LEVEL where you can review your edited values before per...

Page 3: ...to local ambient atmospheric pressure and to compensate for drift To use this function properly the device must be mounted in the application and the pressure sensor exposed to local ambient atmosphe...

Page 4: ...e simulation mode is started No The device remains in its normal operating mode The default setting is No WARNING After you have completed testing in the simulation mode do not forget to return to thi...

Page 5: ...250 mA the load is disconnected and the WARN output is set to 0 5 V in a loop Table 2 DCM Switch functions and definitions definition function equa tion active inactive max monitor N O 1 closed open m...

Page 6: ...ausibility of your values and if necessary redirect you back to the PARAMETRIZATION LEVEL and The CONFIRMATION LEVEL where you can review your edited values before permanently storing them START SCREE...

Page 7: ...2 Wire Trans mitter is 4 20 mA However it can be inverted to 20 4 mA see section Function below FUNCTION Here you can invert the device s function see also Fig 15 Non invert the mode current range or...

Page 8: ...e contrast is increased to its maximum value The default setting is Mean EXTRAS SIMUL SIMULATION MODE The simulation mode allows the user to test the device within the application without actually app...

Page 9: ...passive 0 mA 3 Wire 0 10 V passive 0 V The output limits of the device s outputs at overpressure and underpressure are as follows Table 4 SN Transmitter output limits at over underpressure device out...

Page 10: ...er The Y axis displays the generated output in the range according to the output definition see Fig 15 on page 7 and Table 5 on page 9 DCM SWITCH Measuring pressures is the safety function of the dire...

Page 11: ...output pin 2 If due to miswiring or electronic failure the signal output cannot reach the desired high failure state it will auto matically go to the low failure state Table 6 SN Transmitter error rep...

Page 12: ...ontact FEMA 6 Device memory check failure R U Hardware reset as described above Contact FEMA 7 Device program flow failure R U If unrecoverable Contact FEMA 8 Environment too hot or electronics failur...

Page 13: ...sionen mit HMI 20 70 C Rel Luftfeuchtigkeit 0 95 nicht kondensierend Mediumtemperatur 30 100 C 16 bar Modelle 40 100 C 16 bar Modelle Lagertemperatur Versionen ohne HMI 40 80 C 16 bar 40 100 C 16 bar...

Page 14: ...ener Einstellwerte PR FBEREICH berpr fung auf korrekte Eingabe und ggfls Zur ckstufung in den PARAMETRISIERUNGS BEREICH und BEST TIGUNGSBEREICH Nochmalige Best tigung der Einstellwerte bevor sie dauer...

Page 15: ...st N O AUTO ZERO Die auto zero Funktion erm glicht es dem Anwender das Ger t entsprechend den lokalen atmosph rischen Druck verh ltnissen einzustellen sowie die Drift auszugleichen Um diese Funktion r...

Page 16: ...das Ger t wieder in Sicherheitsanwendungen eingesetzt werden Yes Der Simulationsmodus wird aufgerufen No Das Ger t bleibt in seiner normalen Betriebsart Die Werkseinstellung ist No VORSICHT Nach Absc...

Page 17: ...annungsversorgung gestellt Falls der Strom immer noch h her als 250 mA ist wird die Last getrennt und der WARN Ausgang auf 0 5 V in eine Schleife gestellt Tabelle 2 DCM Funktionen und Definitionen Def...

Page 18: ...ung auf korrekte Eingabe und ggfls Zur ckstufung in den PARAMETRISIERUNGS BEREICH und BEST TIGUNGSBEREICH Nochmalige Best tigung der Einstellwerte bevor sie dauerhaft gespeichert werden STARTANZEIGE B...

Page 19: ...tiert werden siehe Abschnitt Function Funktion unten FUNCTION FUNKTION Hier k nnen Sie die Ger tefunktion invertieren siehe auch Abb 15 Non invert der Modus Strom bzw Spannungsbereich wird auf 0 4 20...

Page 20: ...Simulationsmodus erm glicht es dem Anwender die reibungslose Funktion des Ger ts zu pr fen ohne das System unter Druck zu setzen Nach Aufrufen des Simulationsmodus sind die Diagnostik funktionen nicht...

Page 21: ...V passiv 0 V Die Ausgangsgrenzen der Ger teausg nge bei ber und Unterdruck sind wie folgt Tabelle 4 SN Transmitter Ausgangsgrenzen bei ber und Unterdruck Ger t Ausgangs signal bereich min Ausgangs gre...

Page 22: ...te besteht darin Dr cke zu messen Je nach dem wie das Ger t eingestellt ist l t es sich mit einer Genauigkeit von 5 als Max Min oder Window W chter Schalter ausgang und WARN Ausgang betreiben Wenn di...

Page 23: ...er automatisch in den niedrigen Fehlerzustand fallen Tabelle 6 SN Transmitter Fehlermeldung Fehlermeldung Modell Aus gangs signal Bereich prim rer hoher Fehler zustand sekund rer niedriger Fehler zust...

Page 24: ...ntaktieren 6 Speicherpr fung fehlgeschlagen B NB Wie oben beschrieben Hardware Reset durchf hren FEMA kontaktieren 7 Programmlauffehler B NB Falls nichtbehebbar FEMA kontaktieren 8 Umgebung zu hei ode...

Reviews: