FEMA ELECTRÓNICA . SERIES M . M60-LC

6

Digits

 

number of digits 

6

led  

7 segments

color 

red or green

digit height 

14 mm

Reading

max. reading 

999999

min. reading 

-199999

decimal point 

configurable X.X.X.X.X.X

overrange  

flash reading

underrange 

flash reading

Load cells

type of cells 

1 mV/V, 2 mV/V, 3 mV/V and others

excitation voltage 

configurable 5 Vdc or 10 Vdc

max. excitation current  140 mA

Vexc. protection 

against shortcircuit

 

(see error at section 

1.15

)

max. terminals voltage  30 Vdc

number of cells* 

1 to 8 load cells (power 5 Vdc)

 

1 to 4 load cells (power 10 Vdc)

*(values  calculated  for  standard  350 Ohms  load  cells.  For 

cells with different impedance, the number is limited by the 

140 mA available current)

Measure

signal ranges 

(see section 1.9.2)

accuracy at 25 ºC 

(see section 1.9.2)

thermal stability 

50 ppm/º

input impedance 

20 MOhm

acquisitions / second 

see 

Table 2

   (and refresh for alarms, analog outputs and bus)

display refresh 

see 

Table 2

step response 

see 

Table 2

 

0 % to 99 % signal

signal terminals

 

plug-in screw terminals (pitch 3.81 mm)

Power

power ‘H’ 

85 to 265 Vac/dc

power ‘L’ 

11 to 60 Vdc and 24/48 Vac

isolation* 

2500 Veff with power ‘H’

*(60 seconds)

 

1500 Veff with power ‘L’

consumption 

<1.5 W only meter 

 

<4.0 W meter with options

power terminals

 

plug-in screw terminals (pitch 5.08 mm)

wire section  

1 to 2.5 mm2 (AWG17 to AWG14)

Configuration

 

front keypad with 3 keys

Front protection

 IP65

Output and control options

  relay, analog, communications, ...

 

(see section 2)

Mechanical

mounting 

panel

connections 

plug-in screw terminal

housing  

ABS, polycarbonate (V0)

weight  

<150 gr.

front size  

96 x 48 mm (1/8 DIN)

panel cut-out 

92 x 44 mm

depth  

91 mm (including terminals)

Temperature

  

operation  

0 to +50 ºC

storage 

20 to +70 ºC

warm-up time 

15 minutes

Acquisitions / sec.

Display refresh

Step response

Mode standard

16 acq. / sec.

16 refresh / sec.

63 mSec.

Mode 50 Hz

50 acq. / sec.

16 refresh / sec.

20 mSec.

Mode 60 Hz

60 acq. / sec.

16 refresh / sec.

17 mSec.

Table 2 - Technical data for the configured acquisition mode

The instrument works by default with a fast acquisition mode 

of 16 acquisitions per second, with a noise rejection optimized 

for 50 and 60 Hz frequencies. Two additional faster acquisition 

modes are available, optimized for noise rejection to a single 

specific frequency of 50 Hz or 60 Hz. 
To optimize the noise rejection only to 50 Hz and / or increase 

the acquisition speed to 50 acquisitions per second, configure 

the ‘

Mode

’ (‘

ModE

’) parameter to ‘

50.hZ

’ value. This selec

-

tion increases the speed to 50 acquisitions per second and 

increases the noise rejection to 50 Hz, although it reduces the 

noise rejection to 60 Hz. Configure the parameter value to ‘

60.

hZ

’ to increase to 60 acquisitions per second and maximum 

rejection to 60 Hz noise, reducing the noise rejection for 50 Hz.
To configure the mode see section 

1.12.7

.

1.9 

Technical specifications

1.9.1 Acquisition modes

Summary of Contents for M Series

Page 1: ...nts tare and auto tare functions on power up functions stabil ity alarms configurable operator menu configurable led brightness Universal AC and DC power Up to 3 optional modules for output and contro...

Page 2: ...lue in mV excitation current provided in mA and real excitation voltage see section 1 12 13 configurable Fast access menu key UP 5 with access to selected functions see section 1 12 13 Multiple displa...

Page 3: ...configuration menu 19 1 2 Index 1 12 16 On power up configuration 19 1 12 17 Tools 20 1 12 18 Access to options configurations menu 20 1 13 Full configuration menu 22 1 14 Factory configuration 25 1 1...

Page 4: ...ntrol options see section 2 the shipment also includes 1 pack of green signal terminals for each output and con trol option installed Figure 1 Set of units label 1 5 Additional information User s manu...

Page 5: ...cells 2 3 4 cells or more con nect the sense wires to the electrical middle point of the power wires of all the cells see section 1 16 4 1 8 2 Connecting the cell to the ground Measuring with load ce...

Page 6: ...G14 Configuration front keypad with 3 keys Front protection IP65 Outputandcontroloptions relay analog communications see section 2 Mechanical mounting panel connections plug in screw terminal housing...

Page 7: ...ad cells For load cells with different impedance calculate the current consumption for each cell and the total must not exceed the maximum current the instrument can pro vide In case of problems with...

Page 8: ...o the instrument and now the reading is 0 Kg The truck enters the loading area and when it leaves it is placed on the weighing system again Now the reading is 1 550 Kg This is the weight of the mate r...

Page 9: ...ress the SQ key for 1 second to ac cess the configuration menu Inside the menu the SQ key acts as an ENTER It enters into the menu option selected and when entering a numerical value it validates the...

Page 10: ...is particular load cell For the empirical configuration you will need access to two weights a low weight as small as possible it can be the cell without weight and a high weight as close as possible t...

Page 11: ...d cell accepts any value between 0 and 999999 The decimal point will be shown in the posi tion configured at the Decimal point dP parameter De fault value is 1000 Example for a 5 Kg cell configure a v...

Page 12: ...462 pounds Within the scale factor configure the multiplier to 220462 and the divider to 100000 The instrument is now configured to read in pounds The Stock units Stck function is provided to count la...

Page 13: ...auto tare function The tare is automatically activated when reading is lower than the activation value and the fluctuation of the reading is lower than the counts defined at the stability band for th...

Page 14: ...d the liquid is still moving inside the tank and this movement is reflected in weight and reading oscillations Alarm 2 is configured as stability alarm and activates when the liquid inside the tank is...

Page 15: ...elay dEL 1 parameter config ure the delay to apply before the alarm is deactivated Delay starts to count once the setpoint is reached plus the hysteresis value Value from 0 0 to 99 9 seconds at the St...

Page 16: ...er with higher values Increasing the strength of the filter slows the reading Value 0 disables the filter the Steps StEP function configures the reading to be done in steps of 1 2 5 10 20 or 50 counts...

Page 17: ...bled the activation is per formed sequentially in the same order as the configuration menu first is the tare then the alarm unlock etc 1 12 11 Rear controls 1 12 12 Front key LE 3 The front key LE 3 c...

Page 18: ...p function The On power up on Pu menu allows to define a series of functions to activate when the instrument restarts after a power loss Functions available are a delay on the activation of measure an...

Page 19: ...he Exc voltage c EXc function allows to visualize the actual value of the excitation voltage measured between terminals sense and sense Value is offered in Vdc the Exc current c MA function allows to...

Page 20: ...the Password func tion select on and enter the code The numerical code is asked when accessing the configu ration menu key SQ Functions configured to be ac cessible through the fast access menu are no...

Page 21: ...FEMA ELECTR NICA SERIES M M60 LC 21 T h i s p a g e b l a n k...

Page 22: ...l conf Load cell parameters Decimal point Sensitivity mV V Nominal weight Excitation voltage Field correction High weight correction Low weight correction Low weight value High weight value Stability...

Page 23: ...Fixed digits Fix digits Memory of maximum Memory of minimum Rear controls Tare Alarmunlock Stock units Key LE Tare Alarmunlock Stock units Key UP fast access Gross weight Setpoint 1 Setpoint 2 Setpoi...

Page 24: ...configuration Option 1 Access to the optional module installed at slot 1 Option 2 Access to the optional module installed at slot 2 Option 3 Access to the optional module installed at slot 3 1 14 Ful...

Page 25: ...ff Locked alarms A LcK off Display dISP Fixed digits FIX d no fixed digits Average filter AVr 0 disabled Steps StEP 1 Left zeros LZEr off Memory of maximum MAX 199999 Memory of minimum MIn 999999 Rear...

Page 26: ...powered from the instrument with power 10Vdc nominal weight 100Kg and 2mV V sensi bility Connect the load cell to the instrument see section 1 8 Configure the theoretical configuration of the load ce...

Page 27: ...nt to measure a mil livolt generated signal at the laboratory there is a special configuration to apply Different from the load cell the millivolt generator is not a differential system and does not n...

Page 28: ...he front filter at corner X and then insert clips A B C and D in this order Risk of electric shock Removing the front cover will grant access to the internal circuits Disconnect the input signal to pr...

Page 29: ...tion required Instrument conforms to CE rules and regulations This instrument is warranted against all manufactur ing defects for a period of 24 months as requested by the European legislation This wa...

Page 30: ...mitter B Collector C Not connected Figure 17 Connections for T1 transistor module The R1 module provides 1 relay output to install on Series M meters up to a maximum of 3 relay outputs in the same met...

Page 31: ...t module 2 4 Module AO The AO module provides 1 isolated analog output configu rable for 4 20 mA or 0 10 Vdc to install on Series M me ters up to a maximum of 3 analog outputs in the same meter Output...

Page 32: ...al_en pdf For additional information http fema es docs 4254_SERIES_M K S_OPTIONS_manual_en pdf Figure 22 Detail of the Modbus RTU communications module Power Opt 2 Opt 1 Opt 3 B A G B A G Signal B A G...

Page 33: ...R4 or R6 module disables the alarm menus AL1 AL2 and AL3 of the instrument Choose special modules R2 R4 or R6 when a higher density of relays is needed Type of relay 3 contacts Com NO NC Max current...

Page 34: ...ut front key pad To configure the instru ment remove the meter from the panel and remove the front filter Internal press buttons for configurationareaccessible Op tionally request the instrument preco...

Page 35: ...1 THM benchtop housing Adapter for DIN rail mount for Series M K and C 4 2 Adapter DRA M Adapter 96x96mm for 96x48mm instruments 4 3 Adapter KA96 Wall mount housing To gether with the KIP protec tor...

Page 36: ......

Reviews: