Fema M Series User Manual Download Page 12

FEMA ELECTRÓNICA . SERIES M . M60-LC

12

1.12.4 Function ‘auto-tare’

The  ‘auto-tare’  function  automatically  activates  the  ‘tare’ 

when the weight is removed from the load cell. The ‘auto-

tare’ configuration has three parameters :

• 

Activation value : the ‘auto-tare’ function activates when 

reading is lower than the defined value.

• 

Stability band: reading must be stable, and its fluctuation 

must be lower than the number of counts defined in this pa

-

rameter.

• 

Stability time : reading must be within the stability band for 

the time defined in this parameter.

When these three parameters are met (the system is ‘without 

weight’ and the reading is ‘stable’) the ‘auto-tare’ function au

-

tomatically activates the ‘tare’.

Example : weighing system with reading from 0.0 to 2500.0 

Kg. When weight is removed from the system, there is al-

ways a variable remnant value : 2.2 Kg, 3.1 Kg, -0,7 Kg, ... 

This remnant value is associated to the specific imperfections 

of each load cell. Also this remnant value takes some time to 

stabilize, approximately 1 second. A manual tare can be ap-

plied each time the load cell is unloaded in order to correct 

this error. The ‘auto-tare’ function will correct it automati-

cally without operator intervention, configuring the follow-

ing parameters.

 

activation value = 5.0

 

stability = 1.0 

 

stability time = 2 seconds

When reading is lower than ‘5.0’, the ‘auto-tare’ system ac-

tivates, and it will analyze the stability of the signal. When 

reading does not change more than ±1.0 counts for a time of 

2 seconds, the tare will automatically activate.

The  ‘auto-tare’  function  is  affected  by  the  ‘

Maximum tare

’ 

(

‘MAX.t’

) parameter. The instrument will not accept the acti

-

vation of the tare when reading is higher than the ‘

Maximum 

tare

’ (

‘MAX.t’

) value.

The ‘

Scale factor

’ (‘

ScL.F

’) configures a fixed multiplier to ap

-

ply to the reading.

Example: a weighing system is configured to read in Kg, but 

the system is going to be shipped to an area where measure 

must  be  in  pounds.  The  relation  between  kilograms  and 

pounds  is:  1  Kg = 2,20462  pounds.  Within  the  scale  factor, 

configure the multiplier to 220462 and the divider to 100000. 

The instrument is now configured to read in ‘pounds’.

The ‘

Stock units

’ (‘

Stck

’) function is provided to count large 

quantities of small units, in situations such as stock invento

-

ry, reception of goods, etc. The operator must configure the 

number of ‘units’ assigned to a weight. The instrument will 

measure the weight but will show the number of ‘units’.
To configure the ‘

Stock units

’ function, weight a known num

-

ber of units. Then introduce the number of units, either from 

the  configuration  menu  (‘Advanced  configuration’ \ ‘Stock 

units’) or from the fast access menu (key ‘UP’ (

5

)).

Example from the configuration menu : place 50 units on the 

load cell, and check that the instrument is weighing correctly. 

Enter the configuration menu, and at the ‘Stock units’ (‘Stck’) 

parameter,  within  the  ‘Advanced  configuration’,  introduce 

‘50’ as the number of units. Save the value (key ‘SQ’ (

<

)) 

and leave the configuration menu (key ‘LE’ (

3

) two times). 

The instrument restart and reads 50 units. Add more units 

and observe that the reading increases proportionally to the 

number of units.
Example from the front key ‘UP’ (

5

) : configure the func-

tion ‘Stock units’ to be accessible from the front key ‘UP’ (

5

)’ 

(fast access menu) (see section 1.12.13) and leave the con-

figuration menu. Place 50 units on the load cell, and check 

that the instrument is weighing correctly. Access the param-

eter ‘Stock units’ through the front key ‘UP’ (

5

), and config-

ure the number of units actually on the load cell (50 units). 

Save the value (key ‘SQ’ (

<

)) and leave the fast access menu 

(key ‘LE’ (

3

)). The instrument reads 50 units. Add more units 

and observe that the reading increases proportionally to the 

number of units

In both cases, setting a value to the ‘

Stock units

’ (‘

Stck

’) pa

-

rameter assigns the value to the actual weight. The actual val

-

ue of units (‘

Stock units

’ (‘

Stck

’) parameter) can be assigned 

to the actual weight by pressing front key ‘LE’ (

3

(see section 

1.12.12)

 and/or activating the rear terminal ‘EK’ 

(see section 

1.12.11)

.

Assign the value ‘0’ to the ‘

Stock units

’ (‘

Stck

’) parameter to 

disable this function and return to normal reading of weight.

1.12.5 

Scale factor

1.12.6 Function ‘stock’

Summary of Contents for M Series

Page 1: ...nts tare and auto tare functions on power up functions stabil ity alarms configurable operator menu configurable led brightness Universal AC and DC power Up to 3 optional modules for output and contro...

Page 2: ...lue in mV excitation current provided in mA and real excitation voltage see section 1 12 13 configurable Fast access menu key UP 5 with access to selected functions see section 1 12 13 Multiple displa...

Page 3: ...configuration menu 19 1 2 Index 1 12 16 On power up configuration 19 1 12 17 Tools 20 1 12 18 Access to options configurations menu 20 1 13 Full configuration menu 22 1 14 Factory configuration 25 1 1...

Page 4: ...ntrol options see section 2 the shipment also includes 1 pack of green signal terminals for each output and con trol option installed Figure 1 Set of units label 1 5 Additional information User s manu...

Page 5: ...cells 2 3 4 cells or more con nect the sense wires to the electrical middle point of the power wires of all the cells see section 1 16 4 1 8 2 Connecting the cell to the ground Measuring with load ce...

Page 6: ...G14 Configuration front keypad with 3 keys Front protection IP65 Outputandcontroloptions relay analog communications see section 2 Mechanical mounting panel connections plug in screw terminal housing...

Page 7: ...ad cells For load cells with different impedance calculate the current consumption for each cell and the total must not exceed the maximum current the instrument can pro vide In case of problems with...

Page 8: ...o the instrument and now the reading is 0 Kg The truck enters the loading area and when it leaves it is placed on the weighing system again Now the reading is 1 550 Kg This is the weight of the mate r...

Page 9: ...ress the SQ key for 1 second to ac cess the configuration menu Inside the menu the SQ key acts as an ENTER It enters into the menu option selected and when entering a numerical value it validates the...

Page 10: ...is particular load cell For the empirical configuration you will need access to two weights a low weight as small as possible it can be the cell without weight and a high weight as close as possible t...

Page 11: ...d cell accepts any value between 0 and 999999 The decimal point will be shown in the posi tion configured at the Decimal point dP parameter De fault value is 1000 Example for a 5 Kg cell configure a v...

Page 12: ...462 pounds Within the scale factor configure the multiplier to 220462 and the divider to 100000 The instrument is now configured to read in pounds The Stock units Stck function is provided to count la...

Page 13: ...auto tare function The tare is automatically activated when reading is lower than the activation value and the fluctuation of the reading is lower than the counts defined at the stability band for th...

Page 14: ...d the liquid is still moving inside the tank and this movement is reflected in weight and reading oscillations Alarm 2 is configured as stability alarm and activates when the liquid inside the tank is...

Page 15: ...elay dEL 1 parameter config ure the delay to apply before the alarm is deactivated Delay starts to count once the setpoint is reached plus the hysteresis value Value from 0 0 to 99 9 seconds at the St...

Page 16: ...er with higher values Increasing the strength of the filter slows the reading Value 0 disables the filter the Steps StEP function configures the reading to be done in steps of 1 2 5 10 20 or 50 counts...

Page 17: ...bled the activation is per formed sequentially in the same order as the configuration menu first is the tare then the alarm unlock etc 1 12 11 Rear controls 1 12 12 Front key LE 3 The front key LE 3 c...

Page 18: ...p function The On power up on Pu menu allows to define a series of functions to activate when the instrument restarts after a power loss Functions available are a delay on the activation of measure an...

Page 19: ...he Exc voltage c EXc function allows to visualize the actual value of the excitation voltage measured between terminals sense and sense Value is offered in Vdc the Exc current c MA function allows to...

Page 20: ...the Password func tion select on and enter the code The numerical code is asked when accessing the configu ration menu key SQ Functions configured to be ac cessible through the fast access menu are no...

Page 21: ...FEMA ELECTR NICA SERIES M M60 LC 21 T h i s p a g e b l a n k...

Page 22: ...l conf Load cell parameters Decimal point Sensitivity mV V Nominal weight Excitation voltage Field correction High weight correction Low weight correction Low weight value High weight value Stability...

Page 23: ...Fixed digits Fix digits Memory of maximum Memory of minimum Rear controls Tare Alarmunlock Stock units Key LE Tare Alarmunlock Stock units Key UP fast access Gross weight Setpoint 1 Setpoint 2 Setpoi...

Page 24: ...configuration Option 1 Access to the optional module installed at slot 1 Option 2 Access to the optional module installed at slot 2 Option 3 Access to the optional module installed at slot 3 1 14 Ful...

Page 25: ...ff Locked alarms A LcK off Display dISP Fixed digits FIX d no fixed digits Average filter AVr 0 disabled Steps StEP 1 Left zeros LZEr off Memory of maximum MAX 199999 Memory of minimum MIn 999999 Rear...

Page 26: ...powered from the instrument with power 10Vdc nominal weight 100Kg and 2mV V sensi bility Connect the load cell to the instrument see section 1 8 Configure the theoretical configuration of the load ce...

Page 27: ...nt to measure a mil livolt generated signal at the laboratory there is a special configuration to apply Different from the load cell the millivolt generator is not a differential system and does not n...

Page 28: ...he front filter at corner X and then insert clips A B C and D in this order Risk of electric shock Removing the front cover will grant access to the internal circuits Disconnect the input signal to pr...

Page 29: ...tion required Instrument conforms to CE rules and regulations This instrument is warranted against all manufactur ing defects for a period of 24 months as requested by the European legislation This wa...

Page 30: ...mitter B Collector C Not connected Figure 17 Connections for T1 transistor module The R1 module provides 1 relay output to install on Series M meters up to a maximum of 3 relay outputs in the same met...

Page 31: ...t module 2 4 Module AO The AO module provides 1 isolated analog output configu rable for 4 20 mA or 0 10 Vdc to install on Series M me ters up to a maximum of 3 analog outputs in the same meter Output...

Page 32: ...al_en pdf For additional information http fema es docs 4254_SERIES_M K S_OPTIONS_manual_en pdf Figure 22 Detail of the Modbus RTU communications module Power Opt 2 Opt 1 Opt 3 B A G B A G Signal B A G...

Page 33: ...R4 or R6 module disables the alarm menus AL1 AL2 and AL3 of the instrument Choose special modules R2 R4 or R6 when a higher density of relays is needed Type of relay 3 contacts Com NO NC Max current...

Page 34: ...ut front key pad To configure the instru ment remove the meter from the panel and remove the front filter Internal press buttons for configurationareaccessible Op tionally request the instrument preco...

Page 35: ...1 THM benchtop housing Adapter for DIN rail mount for Series M K and C 4 2 Adapter DRA M Adapter 96x96mm for 96x48mm instruments 4 3 Adapter KA96 Wall mount housing To gether with the KIP protec tor...

Page 36: ......

Reviews: