FEMA ELECTRÓNICA . SERIES M . M60-LC
19
At the ‘
Key UP (‘fast access’)
’ (‘
K.uP
’) menu configure which
functions and parameters will be accessible through the ‘fast
access’ menu. Select ‘on’ to activate each function. See sec
-
tion
for additional information.
•
the ‘
Setpoint
1
’ (‘
ALr1
’)
function allows to visualize and
modify the setpoint for alarm 1.
•
the ‘
Setpoint
2
’ (‘
ALr2
’)
function allows to visualize and
modify the setpoint for alarm 2.
•
the ‘
Setpoint
3
’ (‘
ALr3
’)
function allows to visualize and
modify the setpoint for alarm 3.
•
the
‘
Stock units
’ (‘
Stck
’) function allows to visualize and
modify the quantity of units defined at the ‘stock units’ pa
-
rameter
.
•
the ‘
Gross weight
’ (‘
GroS
’)
function allows to visualize the
gross weight.
•
the ‘
Tare value
’ (‘
tArE
’)
function allows to visualize the ac
-
tual tare value.
•
the ‘
Signal mV
’ (‘
c.MV
’)
function allows to visualize the
actual value of the input signal, without scaling. Value is of
-
fered in mV.
•
the ‘
Exc. voltage
’ (‘
c.EXc
’)
function allows to visualize the
actual value of the excitation voltage, measured between
terminals ‘
sense+
’ and ‘
sense-
’. Value is offered in Vdc.
•
the ‘
Exc. current
’ (‘
c.MA
’)
function allows to visualize the
actual value of the current provided by the instrument to the
load cell. Value is offered in mA.
•
the ‘
Memory of maximum
’ (‘
MAX
’)
or
‘
Memory of mini
-
mum
’ (‘
MIn
’) allows to visualize and/or reset the actual val
-
ue of the maximum and minimum memory.
1.12.15 ‘Fast access’ configuration menu
Key UP
(‘Fast access’)
Gross weight
Setpoint 1
Setpoint 2
Setpoint 3
Tare value
Stock units
Memory of
maximum
Memory of
minimum
Signal mV.
Exc. voltage
Exc. current
The ‘
On power up
’ (‘
on.Pu
’) menu assigns functions to apply
when the instrument restarts after a power loss. See section
for additional information.
•
at the ‘
Delay
’ (‘
dLAy
’)
parameter configure the time the
instrument waits before starting normal operation. Value be
-
tween 0 and 200 seconds.
•
at the ‘
tare
’ (‘
tArE
’)
parameter configure to ‘
on
’ to activate
a tare every time the instrument restarts after a power loss.
1.12.16 ‘On power up’ configuration
On power up
Delay
Tare
Seconds