Fema M Series User Manual Download Page 10

FEMA ELECTRÓNICA . SERIES M . M60-LC

10

1.12.2 Initial set-up

Before starting to configure the instrument, identify the pa

-

rameters of the load cell, at the manufacturers datashee

(see 

Table 5)

If the parameters are not know, leave the instrument 

with the default values.

Load cell  

parameters

Default values  

Sensitivity

2

mV/V

Nominal weight

1000

Kilos

Excitation voltage

10

Vdc

Table 5 - Parameters of the load cell 

For an accurate measure, the instrument needs to correctly 

configure its parameters for the particular load cell connect

-

ed. The configuration procedure has a first theoretical step 

and a second empirical step. The third and final step will set 

the ‘system zero’ of the instrument.

Theoretical configuration of the load cell

The theoretical parameters are configured at the ‘

Parameters 

of the cell

’ (‘

cELL

’) menu. 

• 

at the ‘

Decimal point

’ (‘

dP

’) parameter, place the decimal 

point according to the resolution you want to see.

• 

at the ‘

Nominal weight

’ (‘

LoAd

’) parameter introduce the 

nominal weight of the load cell. The value is entered with the 

resolution configured in the parameter above. 

• 

at the ‘

Sensitivity

’ (‘

MV.V

’) parameter, introduce the value 

of the cell sensitivity.

• 

at the ‘

Excitation voltage

’ (‘

V.EXc

’) parameter, select 5 or 

10 Vdc. (The ‘LAb’ value enables the laboratory mode, for di

-

rect measure from a millivolt generator instead of a load cell 

(see section 1.16.5)).

Example : load cell with 1.95 mV/V sensitivity and a nominal 

value of 5

Kg and power 5 Vdc. To read in grams with a deci-

mal point, configure the theoretical parameters as indicated 

below :

Decimal point : XXXXX.X

 

Sensitivity : 1.95 mV/V

 

Nominal weight : 5000.0

 

Excitation voltage : 5 Vdc

When the theoretical values are configured, leave the config

-

uration menu. Apply a ‘

system zero’

. Force a tare, and place 

different weights to check if the reading is correct. If it is not 

correct, apply the empirical configuration and again the ‘

sys

-

tem zero

’.

Empirical configuration of the load cell

The second part of the load cell configuration is an empiri

-

cal process of field correction. The instrument will detect and 

correct the individual deviations of this particular load cell. 
For the empirical configuration you will need access to two 

weights : a low weight, as small as possible (it can be the cell 

without weight) and a high weight as close as possible to the 

nominal weight of the cell.

In each case the meter will be informed of the real weight 

applied to the cell in order to correct and compensate for the 

measured deviations at the signal. Both corrections are need 

(high and low) for a correct configuration of the load cell.

• 

low weight correction : place the load cell without weight 

or with the smallest weight possible, and access the ‘

Low 

weight correction

’ (‘

F.Lo

’) menu. Press key SQ (‘

<

’), intro

-

duce the value of the weight and press again SQ (‘

<

’). The 

instrument will flash shortly and return to the menu entry 

Low weight correction

’ (‘

F.Lo

’).

• 

high weight correction : place the load cell with a weight 

closest to nominal and access the ‘

High weight correction

’ 

(‘

F.hI

’) menu. Press key SQ (‘

<

’), introduce the value of the 

weight and press again SQ (‘

<

’). The instrument will flash 

shortly and return to the menu entry ‘

High weight correc-

tion

’ (‘

F.hI

’).

Once  both  corrections  are  applied,  leave  the  configuration 

menu. Force a tare, and place different weights to check that 

the reading is correct. As a last step, assign the ‘system zero’ if 

you want to access gross weight and net values.

Assign the ‘system zero’

This is a necessary and important step for a correct measure

-

ment with a load cell.

• 

assign the ‘system zero’ : place the load cell without weight 

or with the weight that will be considered as ‘zero’ and access 

the ‘

System zero

’ (‘

S.ZEr

’) parameter. Press key SQ (‘

<

’). The 

instrument will flash shortly and return to the menu entry 

System zero

’ (‘

S.ZEr

’).

The empirical configuration of the load cell recalcu

-

lates  and  updates  the  theoretical  sensitivity  value 

(‘

Sensitivity    mV/V

’  (‘

MV.V

’)  parameter).  Manual 

modifications of this parameter will modify the configura

-

tion  of  the  cell.  To  prevent  accidental  modification  con

-

sider the activation of the ‘

password

’ function 

(see section 

1.12.17)

.

!

Once  the  load  cell  has  been  correctly  configured, 

and the reading of the instrument is correct, it is not 

necessary to access again this part of the configura

-

tion menu. If you need to scale the reading to different units, 

use the ‘

Scale factor

’ (‘

ScL.F

) parameters at the ‘Advanced 

configuration’ menu 

(see section 1.12.7)

.

!

Summary of Contents for M Series

Page 1: ...nts tare and auto tare functions on power up functions stabil ity alarms configurable operator menu configurable led brightness Universal AC and DC power Up to 3 optional modules for output and contro...

Page 2: ...lue in mV excitation current provided in mA and real excitation voltage see section 1 12 13 configurable Fast access menu key UP 5 with access to selected functions see section 1 12 13 Multiple displa...

Page 3: ...configuration menu 19 1 2 Index 1 12 16 On power up configuration 19 1 12 17 Tools 20 1 12 18 Access to options configurations menu 20 1 13 Full configuration menu 22 1 14 Factory configuration 25 1 1...

Page 4: ...ntrol options see section 2 the shipment also includes 1 pack of green signal terminals for each output and con trol option installed Figure 1 Set of units label 1 5 Additional information User s manu...

Page 5: ...cells 2 3 4 cells or more con nect the sense wires to the electrical middle point of the power wires of all the cells see section 1 16 4 1 8 2 Connecting the cell to the ground Measuring with load ce...

Page 6: ...G14 Configuration front keypad with 3 keys Front protection IP65 Outputandcontroloptions relay analog communications see section 2 Mechanical mounting panel connections plug in screw terminal housing...

Page 7: ...ad cells For load cells with different impedance calculate the current consumption for each cell and the total must not exceed the maximum current the instrument can pro vide In case of problems with...

Page 8: ...o the instrument and now the reading is 0 Kg The truck enters the loading area and when it leaves it is placed on the weighing system again Now the reading is 1 550 Kg This is the weight of the mate r...

Page 9: ...ress the SQ key for 1 second to ac cess the configuration menu Inside the menu the SQ key acts as an ENTER It enters into the menu option selected and when entering a numerical value it validates the...

Page 10: ...is particular load cell For the empirical configuration you will need access to two weights a low weight as small as possible it can be the cell without weight and a high weight as close as possible t...

Page 11: ...d cell accepts any value between 0 and 999999 The decimal point will be shown in the posi tion configured at the Decimal point dP parameter De fault value is 1000 Example for a 5 Kg cell configure a v...

Page 12: ...462 pounds Within the scale factor configure the multiplier to 220462 and the divider to 100000 The instrument is now configured to read in pounds The Stock units Stck function is provided to count la...

Page 13: ...auto tare function The tare is automatically activated when reading is lower than the activation value and the fluctuation of the reading is lower than the counts defined at the stability band for th...

Page 14: ...d the liquid is still moving inside the tank and this movement is reflected in weight and reading oscillations Alarm 2 is configured as stability alarm and activates when the liquid inside the tank is...

Page 15: ...elay dEL 1 parameter config ure the delay to apply before the alarm is deactivated Delay starts to count once the setpoint is reached plus the hysteresis value Value from 0 0 to 99 9 seconds at the St...

Page 16: ...er with higher values Increasing the strength of the filter slows the reading Value 0 disables the filter the Steps StEP function configures the reading to be done in steps of 1 2 5 10 20 or 50 counts...

Page 17: ...bled the activation is per formed sequentially in the same order as the configuration menu first is the tare then the alarm unlock etc 1 12 11 Rear controls 1 12 12 Front key LE 3 The front key LE 3 c...

Page 18: ...p function The On power up on Pu menu allows to define a series of functions to activate when the instrument restarts after a power loss Functions available are a delay on the activation of measure an...

Page 19: ...he Exc voltage c EXc function allows to visualize the actual value of the excitation voltage measured between terminals sense and sense Value is offered in Vdc the Exc current c MA function allows to...

Page 20: ...the Password func tion select on and enter the code The numerical code is asked when accessing the configu ration menu key SQ Functions configured to be ac cessible through the fast access menu are no...

Page 21: ...FEMA ELECTR NICA SERIES M M60 LC 21 T h i s p a g e b l a n k...

Page 22: ...l conf Load cell parameters Decimal point Sensitivity mV V Nominal weight Excitation voltage Field correction High weight correction Low weight correction Low weight value High weight value Stability...

Page 23: ...Fixed digits Fix digits Memory of maximum Memory of minimum Rear controls Tare Alarmunlock Stock units Key LE Tare Alarmunlock Stock units Key UP fast access Gross weight Setpoint 1 Setpoint 2 Setpoi...

Page 24: ...configuration Option 1 Access to the optional module installed at slot 1 Option 2 Access to the optional module installed at slot 2 Option 3 Access to the optional module installed at slot 3 1 14 Ful...

Page 25: ...ff Locked alarms A LcK off Display dISP Fixed digits FIX d no fixed digits Average filter AVr 0 disabled Steps StEP 1 Left zeros LZEr off Memory of maximum MAX 199999 Memory of minimum MIn 999999 Rear...

Page 26: ...powered from the instrument with power 10Vdc nominal weight 100Kg and 2mV V sensi bility Connect the load cell to the instrument see section 1 8 Configure the theoretical configuration of the load ce...

Page 27: ...nt to measure a mil livolt generated signal at the laboratory there is a special configuration to apply Different from the load cell the millivolt generator is not a differential system and does not n...

Page 28: ...he front filter at corner X and then insert clips A B C and D in this order Risk of electric shock Removing the front cover will grant access to the internal circuits Disconnect the input signal to pr...

Page 29: ...tion required Instrument conforms to CE rules and regulations This instrument is warranted against all manufactur ing defects for a period of 24 months as requested by the European legislation This wa...

Page 30: ...mitter B Collector C Not connected Figure 17 Connections for T1 transistor module The R1 module provides 1 relay output to install on Series M meters up to a maximum of 3 relay outputs in the same met...

Page 31: ...t module 2 4 Module AO The AO module provides 1 isolated analog output configu rable for 4 20 mA or 0 10 Vdc to install on Series M me ters up to a maximum of 3 analog outputs in the same meter Output...

Page 32: ...al_en pdf For additional information http fema es docs 4254_SERIES_M K S_OPTIONS_manual_en pdf Figure 22 Detail of the Modbus RTU communications module Power Opt 2 Opt 1 Opt 3 B A G B A G Signal B A G...

Page 33: ...R4 or R6 module disables the alarm menus AL1 AL2 and AL3 of the instrument Choose special modules R2 R4 or R6 when a higher density of relays is needed Type of relay 3 contacts Com NO NC Max current...

Page 34: ...ut front key pad To configure the instru ment remove the meter from the panel and remove the front filter Internal press buttons for configurationareaccessible Op tionally request the instrument preco...

Page 35: ...1 THM benchtop housing Adapter for DIN rail mount for Series M K and C 4 2 Adapter DRA M Adapter 96x96mm for 96x48mm instruments 4 3 Adapter KA96 Wall mount housing To gether with the KIP protec tor...

Page 36: ......

Reviews: