FEMA ELECTRÓNICA . Series K . K40-C1
3
1.1 How to order
K40
C1
model
Power
-
Option 1
Others
H
-
-
-
Option 2
-
Option 3
-
-r1
(1 relay)
-AO
(analog output)
-RTU
(Modbus RTU)
-S4
(RS-485)
-S2
(RS-232)
-
(empty)
-H
(85-265
Vac/dc)
-L
(11/60
Vdc,
24
Vac, 48
Vac)
-NBt
(no buttons)
-65
(front IP65)
1.2 Impulse counter modes
The instrument allows for 5 selectable impulse counter modes :
• Counter (‘
cn.1
• Counter quadrature (‘
cnq.2
• Counter with inhibition (‘
cnI.3
• Counter with control add / substract (‘
cnc.4
• Counter differential (‘
cnd.5
Configurable up or down counting, ‘reset’ at rear terminals, front key
and/or at alarm activation. Configurable ‘Preset’ value. Relay activa-
tion and deactivation delays.
Alarm functions with ‘return to preset’ or ‘reset to 0’ generate cycles
of counting (instrument counts from ‘preset’ value to alarm value
in never ending cycle). The instrument provides memory of cycles
counter.
Scalable reading with configurable multiplier factor (1 to 9999) and
configurable divider factor (1 to 9999). Memory retention in case or
power loss. Retains configuration and last reading.
Counting frequency up to 250 KHz, in ‘
FAST
’ mode (see section 1.16).
1.3 Ratemeter modes
In ratemeter mode the reading is proportional to the measured fre-
quency. The instrument allows for 2 selectable ratemeter modes :
• Ratemeter (‘
rt.6
• Ratemeter quadrature (‘
rtq.7
The ratemeter mode has a single input channel, with scalable read-
ing. The quadrature ratemeter mode has 2 inpur channels available
for detection of sense of turn when working with quadrature signals.
Scalable reading with configurable multiplier factor (1 to 9999) and
configurable divider factor (1 to 9999).
For low frequency applications, the ‘
SLOW
’ mode provides the best
response time for each application (see section 1.15).
Maximum frequency up to 900 KHz and minimum frequency down to
de 0.001 Hz (1 mHz) with ‘
SLOW
’ mode active.
Reading is proportional to the signal period. Scalable reading with
configurable multiplier factor (1 to 9999) and configurable divider
factor (1 to 9999).
For applications with long periods (slow frequencies), the ‘
SLOW
’
mode provides the best response time for each application
1.4 Periodmeter mode
Functions included
Section
‘Fast access’
yes
‘SLOW’ mode
yes, for slow frequencies
‘FAST’ mode
yes, for fast counting
Multiplier and divider
from 1 to 9999
Reset configurable
yes (front, rear and linked
to alarm activation)
Preset
yes
Trigger level
configurable
‘Trigger Sense’ function
helps to set the trigger
Sensor selection
by menu
Cycle counter
Retention memory
yes
, recovers with power
‘On Power Up’
yes
Excitation voltage
configurable
Display filters
recursive
Memory
max., min., cycles
Password
blocks access to configu-
ration menu
Alarms
double setpoints
activation delays
deactivation delays
hysteresis
inverted relays
locked alarms
Display brightness
5 levels
Table 1 - Functions included
1.5
Functions included