F-940 Instruction Manual rev. 11/26/2019
1554 NE 3
rd
Ave, Camas, WA 98607, USA
56
Phone: (360) 833-8835
www.felixinstruments.com
Please note the ranges above. A sensor may
pass this diagnostic test, but still not fall within
the reported accuracy range of the specified
sensor. This means that even if a sensor
passes this diagnostic, it may need calibration
depending on the accuracy desired.
The final automatic diagnostic test is Checking
Calibration Parameters. This diagnostic test
checks the gain and offset values of each
sensor and verifies that they fall within a typical
range. If an error is detected the reported
recommendation will be to zero, calibrate, or
reset factory defaults.
If you are prompted to run hardware
diagnostics by a Felix Support Specialist or
technician, copy the results of the diagnostic
tests by using the clipboard icon at the bottom
right of the screen and send these results to
your correspondent.
Copy
diagnostic
test results