to Setup > Mode > Measure >
Continuous
c. In main menu, navigate to
Measure
, then press the start
button.
d. Put the instrument up to your
ear, with your ear on the Felix
logo on the front, listen for the
sound of the pump. It should be
continuously running
e. With your finger block the Inlet
port (port on the right as you are
looking at the instrument), within
1-3 seconds you should hear
the sound of the pump clicking,
immediately release your finger
from the Inlet port. If you hear
the clicking sound proceed to
step f, if you do not proceed to
the Leak Test Fail Instructions
on page 91.
f. With your finger block the Outlet
port, within 1-3 seconds you