Step-by-Step Assembly:
Page 7
Condenser Assembly:
Rotate the coarse focusing knob (1) to raise the stage
to its highest setting.
Rotate the condenser height adjustment knob (2) to
lower the condenser bracket to its lowest setting.
Loosen the condenser set screw (3).
Swing out the front lens of the condenser with the
scale facing forward. Line up the condenser screw (4)
with the groove (5) of the condenser holder.
Tighten the condenser set screw (3) and raise the
condenser to the highest position with the
condenser height adjustment knob.
Nosepiece Assembly:
Loosen the set screw (1) on the microscope arm.
Line up the dovetail interface (2) of the nosepiece
with the dovetail groove on the microscope arm and
insert it into the body.
Tighten the set screw (1) on the microscope arm.
Head Assembly:
Loosen the head set screw (1) on the microscope
arm.
From a right angled position, insert the coattail
interface on the bottom of the head into the hole in
the middle of the arm. Keep the eyetubes inclined
forward.
Tighten the set screw (1) on the microscope arm.
Objective Assembly:
Rotate the coarse focusing knob to lower the stage.
Install the objectives (1) into the nosepiece from the
lowest magnification to the highest magnification in
a clockwise direction.
Search and focus for a sample starting with the
lowest objective (4x or 10x) when operating. Then
move up to a higher magnification.