Fluorescence Assembly:
Page 15
Nosepiece Assembly:
Loosen the set screw on the arm (1).
Match the dovetail interface (2) of the nosepiece
with the dovetail groove of the arm. Push the
nosepiece into the slot.
Tighten the set screw (1).
Fluorescence Light Source Assembly:
Loosen the set screw (1) on the fluorescence
illuminator.
Push the light source holder (2) into the fluorescence
illuminator holder (3). Make sure the upper plane of
the light source group is horizontal.
Tighten the set srew (1).
When operating, make sure there is enough space
around the light source for heat to escape, especially
on the top and bottom.
Head Assembly:
Loosen the set screw (1) on the fluorescence
illuminator.
From the right side, insert the dovetail interface on
the bottom of the head into the hole in the middle
of the head with a slight incline. Keep the eyetubes
facing forward.
Tighten the set screw (1).
Assemble the eyepieces and objectives.