FEASA LED ANALYSER
ICT VERSION
Capture Sequence Mode
Set/Put SequenceMode# - Set sequence mode for Fibers
Transmit
Receive
setsequencemode#
putsequencemode#
OK
OK
Where:
#
represents the sequence mode and is a number in the range 0 – 1.
Description
This command is used to set the number of fibers that will be used in the capture sequence
command. PutSequencemode
0
will set
all
Fiber Channels for the Capture Sequence.
PutSequencemode
1
sets a
single
Fiber Channel for the capture sequence.
If
put
s
equencemode
1
is selected the Fiber Channel must be set using the
putSequenceChannel
command. Also, when this mode is selected up to 8000 captures can
be stored in memory.
If putsequencemode
0
is selected the number of captures is limited to 400.
putsequencemode
commands are stored in RAM memory while
setsequencemode
commands are stored in Flash memory.
Example 1:
The PC transmits
putsequencemode0
to the LED Analyser to instruct it to do sequence capture
on all Fibers. The LED Analyser will return a string
OK
to the PC.
putsequencemode0
OK
Example 2:
The PC transmits
putsequencemode1
to the LED Analyser followed by
putsequencechannel05
to instruct it to do sequence capture on Fiber No. 5. The LED Analyser will return a string
OK
to the PC.
putsequencemode1
OK
putsequencechannel05
OK
To determine which sequence mode is set, use the command
getsequencemode
.
It will
return a 1 or a 0.
getsequencemode
0
(All Fibers)
111