FEASA LED ANALYSER
ICT VERSION
Set Data Mode
setFREQUENCYxxx - Set the Frequency Output Mode of the Analyser
Transmit
Receive
SetFrequencyxxx
OK
Where:
xxx represents the Frequency Measuring mode of the Analyser at ICT
HSI
represents the Hue, Saturation & Intensity mode.
XyI
represents the Xchromaticity, yChromaticity & Intensity mode.
WSI
represents the Dominant Wavelength, Saturation & Intensity mode.
Description
The Analyser shipped from the factory is defaulted to FrequencyOut-HSI (Hue, Saturation and
Intensity) Mode, There are 2 other frequency out modes available – WSI (Dominant
Wavelength, Saturation, Intensity) and xyI (xy chromaticity and Intensity)
These commands are used to set the Analyser ICT Interface in
Frequency Output
mode of your
choice. It must be issued through the Serial Port. The unit will remain in this mode until
instructed otherwise. The status of the ICT Interface can be determined using the
command.
Example:
The PC transmits
setfrequencyhsi
to the LED Analyser to instruct it to set Frequency Output
Hue, Saturation & Intensity mode.
The PC transmits
setfrequencywsi
to the LED Analyser to instruct it to set Frequency Output
Wavelength, Saturation & Intensity mode.
The PC transmits
setfrequencyxyi
to the LED Analyser to instruct it to set Frequency Output
xChromaticity, yChromaticity & Intensity mode.
Setfrequencyhsi
OK
Setfrequencywsi
OK
Setfrequencyxyi
OK
95