FEASA LED ANALYSER
ICT VERSION
Synchronous Serial Mode
Capture for PWM LED's
Pulse-Width-Modulated(PWM) LED's are switched on and off rapidly to save power and to
control Intensity. The Analyser contains a special set of commands for testing PWM LED's.
There are five Intensity ranges that can be specified to get the best results from the test.
The ranges are designated Range1, Range2, Range3, Range4 and Range5. Range1 has the
highest sensitivity for low light conditions. The sensitivity is reduced progressively to Range5
which has the lowest sensitivity. This means that for brighter LED's the higher Ranges should
be used.
The command codes are shown in Figure 25.
Command
Data
Transmitted
Data Received
Cspture Auto Mode
0xB0 + byte2
0xD5 or 0xEE (error condition)
Capture using Range 1
0xB1 + byte2
0xD5 or 0xEE (error condition)
Capture using Range 2
0xB2 + byte2
0xD5 or 0xEE (error condition)
Capture using Range 3
0xB3 + byte2
0xD5 or 0xEE (error condition)
Capture using Range 4
0xB4 + byte2
0xD5 or 0xEE (error condition)
Capture using Range 5
0xB5 + byte2
0xD5 or 0xEE (error condition)
Figure 25.
The data in byte2 is a hexadecimal number in the range 0x01 – 0x0F. This number represents
the
Averaging Factor
for PWM testing. The smaller the value the faster the test will be to
execute. However, this value should be derived experimentally to determine the best
compromise between speed and stability.
It is recommended to use an initial value of 0x07.