DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 23 of 63
2.3. TURN OFF THE EQUIPMENT
To turn the equipment off, turn the selector switch to the OFF position (see next figure).
2.4. INTERNAL BATTERY
2.4.1. WARNING MESSAGES
WARNINGS
•
The battery must be replaced by an original battery for this equipment. Replacement with another type of battery can result in
an unacceptable risk of temperature rise, fire and/or explosion of the battery or the equipment.
CAUTIONS
•
The equipment has rechargeable battery type must remain connected to the power line during periods when it is not used.
•
If the capacity of the internal battery is below 80% the equipment will be able to function from line power or from External
+12Vdc.
•
Do not discharge the battery completely.
•
Recharge the battery immediately after it’s been used.
•
The internal battery of this equipment cannot be replaced by the user. It has to be replaced by qualified and authorized
personnel.
•
In case you replace the battery follow the local instructions to dispose Ni-Mh batteries or send them to
feas
ELECTRÓNICA
for disposal
.
•
When the device is stored in warehouse, should be put to charge at least once every 60 days for at least 3hrs at 25ºC ±3ºC
temperature, to prevent battery damage.
2.4.2. DESCRIPTION
The device has rechargeable battery type must remain connected to the power line during periods when it is not used.
In case of disconnection or failure of the Power Line or external 12Vdc, the equipment may still be operated via the internal battery.
If, in addition, the internal battery is discharged or damaged, the unit will shut down. It will remain off until the operator turns it back on
(move the power selector switch to any position except OFF) by restoring the power. The Power Line presence is indicated on the front
of the equipment with a green light on the
symbol. When you reconnect the equipment to the Power Line or external 12Vdc (no
matter the elapsed time), will remain off until you turn on the device. Moreover, the device will turn on with the same setup which was
shut down with, except for the minimum alarm limit SpO
2
, which if it was off or set to less than 85%, will be adjusted to 85%, by
regulatory requirement. If there been set to more than 85%, it retains the value set by the operator.
When connected to the power line or external 12Vdc, is charging the internal battery, whether the device is off or on. This condition is
expressed in the green light over the
symbol.
There is no need disconnecting and reconnecting the device periodically. This device has an automatic charging system that allows you
to keep the battery at full charge state.
To power the device with external 12Vdc (ambulance or other vehicle), you have to use the cable with lighter plug. In this condition, it
also recharges the battery and is expressed in the green light over the
symbol.
The state of battery charge is indicated on the screen and, therefore, requires the device have to be switched on (in MONITOR mode or
selection of energy), and operating from the internal battery (disconnected from AC line and from external 12Vdc). When battery is fully