DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 61 of 63
Electromagnetic Immunity
The
feas
ELECTRÓNICA
’s DEFIBRILLATOR Monitor Mod. 3850B-Biphasic REC/TPM/SpO
2
is intended for use in the
electromagnetic environment specified below. The customer or the user of the Defibrillator should assure that it is
used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the
feas
ELECTRÓNICA
’s DEFIBRILLATOR
Monitor Mod. 3850B-Biphasic REC/TPM/SpO
2
,
including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3Vrms
150kHz to 80MHz
3V
1.17 √
10Vrms
150kHz to 80MHz in
ISM bands
a
10V
1.2 √
Radiated RF
IEC 61000-4-3
10V/m
of 80MHz to 2.5GHz
10V/m
1.2 √
80MHz to 800MHz
2.3 √
800MHz to 2.5GHz
where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in meters
(m)
b
.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
c
,
should be less than the compliance level in each
frequency range
d
.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note 1:
At 80MHz and 800MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
The ISM (industrial, scientific, and medical) bands between 150kHz and 80MHz are 6.765MHz to 6.795MHz, 13.553MHz to
13.567MHz, 26.957MHz to 27.283MHz and 40.66MHz to 40.70MHz.
b
The compliance levels in the ISM frequency bands between 150kHz and 80MHz and in the frequency range 80MHz to 2.5GHz are
intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently
brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended separation distance
for transmitters in these frequency ranges.
c
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the
feas
ELECTRÓNICA
’s DEFIBRILLATOR Monitor Mod. 3850B-Biphasic REC/TPM/SpO
2
is
used exceeds the applicable RF compliance level above, the
feas
ELECTRÓNICA
’s DEFIBRILLATOR Monitor Mod. 3850B-Biphasic
REC/TPM/SpO
2
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating
feas
ELECTRÓNICA
’s DEFIBRILLATOR Monitor Mod. 3850B-Biphasic REC/TPM/SpO
2
.
d
Over the frequency range 150kHz to 80MHz, field strengths should be less than 10V/m.