FAURE-HERMAN
COMPONENT MAINTENANCE MANUAL
744-082 / 744-082-3
28-49-91
PAGE 303/304
JULY 17/13
-
To remove an edge-lighting lamp (540), remove filter (530) and the lamp and retain sleeve
(550).
G.
Removing receptacle connector
(1)
Flowmeter 744-082
-
Mark and unsolder the connecting wires.
-
Remove four attaching screws (310) and retain washers (320), nuts (330) and terminal lug
(340).
-
Remove the connector (300).
(2)
Flowmeter 744-082-3 (See IPL, figure 7)
-
Mark and unsolder the connecting wires and components
-
Remove four attaching screws (70) (80), retain washers (90), terminals (110), nuts (100),
lug (120), bar (130) and square (140).
-
Remove the connector (60).
The document reference is online, please check the correspondence between the online documentation and the printed version.