7.F HEPA FILTERS (by technical assistance personnel)
ATTENTION: before replacing HEPA filters, the isolator must be decontaminated (see Para. 7C.) and
a sterilization certificate must be issued to the technicians before starting the service
operation. For the safety of the personnel and the environment, the use of relevant
personal protection devices is recommended as well as the collection of the replaced
HEPA filters in polyethylene bags.
Pass box top HEPA filters (F3)
1.
Remove the pre-filter support frame.
2.
Unscrew the bolts [20] that fasten the top pass box filter [21]
3.
Replace the clogged filter
4.
Repeat steps from 3 to 1.
Pass box rear HEPA filters (F4) (see diagram for maintenance operations page 44)
5.
Open the external door of the pass box
6.
Unscrew the two screws [7] that fasten the rear grid [22] and remove it
7.
Unscrew the screws of the frame to release the HEPA filter [20]
8.
Repleace the HEPA filter
9.
Repeat the steps from 8 to 5.
20
21
7
22
Summary of Contents for ASEPTIC
Page 45: ...a D GROUP Company 00254EN Rev 01 11 2012 45 13 IDIAGRAM FOR MAINTENANCE OPERATIONS...
Page 46: ...a D GROUP Company 00254EN Rev 01 11 2012 46...
Page 47: ...a D GROUP Company 00254EN Rev 01 11 2012 47 13 J FRONTAL DIAGRAM...
Page 48: ...a D GROUP Company 00254EN Rev 01 11 2012 48 13 K SIDE DIAGRAM...
Page 49: ...a D GROUP Company 00254EN Rev 01 11 2012 49 14 SUPPORT TABLE...
Page 52: ...a D GROUP Company 00254EN Rev 01 11 2012 52 17 WIRING DIAGRAM ASEPTIC EQUIPPED WITH CLARUS L 2...
Page 53: ...a D GROUP Company 00254EN Rev 01 11 2012 53...
Page 54: ...a D GROUP Company 00254EN Rev 01 11 2012 54...
Page 55: ...a D GROUP Company 00254EN Rev 01 11 2012 55...
Page 56: ...a D GROUP Company 00254EN Rev 01 11 2012 56...
Page 57: ...a D GROUP Company 00254EN Rev 01 11 2012 57...
Page 58: ...a D GROUP Company 00254EN Rev 01 11 2012 58...
Page 59: ...a D GROUP Company 00254EN Rev 01 11 2012 59...
Page 60: ...a D GROUP Company 00254EN Rev 01 11 2012 60 18 WIRING DIAGRAM ASEPTIC EQUIPPED WITH STERIS VHP...
Page 61: ...a D GROUP Company 00254EN Rev 01 11 2012 61...
Page 62: ...a D GROUP Company 00254EN Rev 01 11 2012 62...
Page 63: ...a D GROUP Company 00254EN Rev 01 11 2012 63...
Page 64: ...a D GROUP Company 00254EN Rev 01 11 2012 64...
Page 65: ...a D GROUP Company 00254EN Rev 01 11 2012 65...
Page 66: ...a D GROUP Company 00254EN Rev 01 11 2012 66...