8 UP/DOWN ARROWS
Use the arrow keys to scroll the menu, to program changing parameters and
to put in the password. Three passwords are programmed: 1) to start the
isolator – 2) to enter the operator menu – 3) to enter the main menu to
change the data input (allowed only to authorized technical staff – service –
because unsuitable interventions can cause troubles and incorrect operation
of the isolator).
9 ESC
ESC key deletes the operation of data input and goes back to the starting
condition.
10 SET
SET key lets you enter the different functions or confirm the data input going
back to the upper level.
11 STATUS
If pressed in sequence, the following data will appear on the display:
External Temperature: Shows the temperature outside of the isolator; the
LCD will display (for example) "EXT. TEMPERATURE=27°C" This value is
taken by an electron probe installed outside the isolator.
Internal Temperature: Shows the internal temperature of the isolator work
area; the LCD will display (for example) "INT. TEMPERATURE =30°C”. This
value is taken by an electron probe located inside the isolator.
U.V. Lamp Residual Lifetime: Shows the operating time of the U.V. lamp
pre-set by the user with the appropriate keys. The LCD will display (for
example) "U.V. TIME=XXXX h”. When such time is over, the message "U.V.
LIFETIME OVER" will appear on the line below.
F1, F2, F3 and F4 Residual lifetimes of filters: it is the operation time of the
filters installed in the isolator that can be programmed by the user.
The LCD will display (for example)” RES. TIME FILTER 1=XXXX:YY h:min”.
When such time is over, the message “CHECK FILTER (es.) 1”. Will appear
on the line below.
The filters installed in the isolator follow the numbering listed below
TYPE of FILTERS
NUMBER
LAF HEPA
1
SUCTION HEPA
2
INLET HEPA (back of the pass boxes)
3
OUTLET HEPA (Top of the pass boxes)
4
Not Applicable
5
LAF Power: it is shown indirectly by the power supply voltage of the main
motor, expressed as percentage of max. load voltage displayed also in
proportion by a bar.
The display shows the notice: "MOT.LAF = XX % " (max.100%).
EXH Power: it is shown indirectly by the power supply voltage of the suction
motor, expressed as percentage of max. load voltage displayed also in
proportion by a bar.
The display shows the notice: "MOT.EXH = XX % " (max.100%).
Summary of Contents for ASEPTIC
Page 45: ...a D GROUP Company 00254EN Rev 01 11 2012 45 13 IDIAGRAM FOR MAINTENANCE OPERATIONS...
Page 46: ...a D GROUP Company 00254EN Rev 01 11 2012 46...
Page 47: ...a D GROUP Company 00254EN Rev 01 11 2012 47 13 J FRONTAL DIAGRAM...
Page 48: ...a D GROUP Company 00254EN Rev 01 11 2012 48 13 K SIDE DIAGRAM...
Page 49: ...a D GROUP Company 00254EN Rev 01 11 2012 49 14 SUPPORT TABLE...
Page 52: ...a D GROUP Company 00254EN Rev 01 11 2012 52 17 WIRING DIAGRAM ASEPTIC EQUIPPED WITH CLARUS L 2...
Page 53: ...a D GROUP Company 00254EN Rev 01 11 2012 53...
Page 54: ...a D GROUP Company 00254EN Rev 01 11 2012 54...
Page 55: ...a D GROUP Company 00254EN Rev 01 11 2012 55...
Page 56: ...a D GROUP Company 00254EN Rev 01 11 2012 56...
Page 57: ...a D GROUP Company 00254EN Rev 01 11 2012 57...
Page 58: ...a D GROUP Company 00254EN Rev 01 11 2012 58...
Page 59: ...a D GROUP Company 00254EN Rev 01 11 2012 59...
Page 60: ...a D GROUP Company 00254EN Rev 01 11 2012 60 18 WIRING DIAGRAM ASEPTIC EQUIPPED WITH STERIS VHP...
Page 61: ...a D GROUP Company 00254EN Rev 01 11 2012 61...
Page 62: ...a D GROUP Company 00254EN Rev 01 11 2012 62...
Page 63: ...a D GROUP Company 00254EN Rev 01 11 2012 63...
Page 64: ...a D GROUP Company 00254EN Rev 01 11 2012 64...
Page 65: ...a D GROUP Company 00254EN Rev 01 11 2012 65...
Page 66: ...a D GROUP Company 00254EN Rev 01 11 2012 66...