13.F CONTROL OF ALARMS
ALARM FOR MISFUNCTIONING OF MAIN MOTOR-FAN (LAF)
This alarm can be simulated by disconnecting the feeding cables of the main motor-fan. When the isolator is
switched on and the “stand-by” period is over, check that the red light of the relevant alarm and the visual and
sound (buzzer) signals start.
ALARM FOR MISFUNCTIONING OF SECONDARY MOTOR-FAN (SUCTION)
Also this alarm can be simulated by disconnecting the feeding cables of the suction motor-fan. When the
isolator is switched on and the “stand-by” period is over, check that the red light of the relevant alarm and the
visual and sound (buzzer) signals start.
ALARM FOR REDUCTION OF THE PRESSURE CAPACITY
To verify this alarm, gradually remove one of the front gloves at the end of stand-by period and check that the
red light of the relevant alarm and the visual and sound (buzzer) signals start. Furthermore check that when
the glove is back in the correct position, the alarm stops.
ALARM FOR OPENING OF THE FRONT WINDOW OF THE ISOLATOR
To verify this alarm it is enough you move the front window from its standard work position removing the four
knobs and check that the red light of the relevant alarm and the visual and sound (buzzer) signals start. Re-
position the window in its standard work position and check that the alarm stops.
ALARM FOR REDUCTION OF LAMINAR AIRFLOW VELOCITY IN THE WORK CHAMBER
During the normal working of the isolator, cover the work surface almost completely with a PVC or paper
sheets and check that the relevant red light switches on. When the correct operating conditions of the laminar
airflow are restored, the alarm will stop.
OTHER ALARMS
With the help of FASTER’s technicians also the alarms for LAF and internal pressure below the minimum
threshold or over the maximum threshold can be checked
Summary of Contents for ASEPTIC
Page 45: ...a D GROUP Company 00254EN Rev 01 11 2012 45 13 IDIAGRAM FOR MAINTENANCE OPERATIONS...
Page 46: ...a D GROUP Company 00254EN Rev 01 11 2012 46...
Page 47: ...a D GROUP Company 00254EN Rev 01 11 2012 47 13 J FRONTAL DIAGRAM...
Page 48: ...a D GROUP Company 00254EN Rev 01 11 2012 48 13 K SIDE DIAGRAM...
Page 49: ...a D GROUP Company 00254EN Rev 01 11 2012 49 14 SUPPORT TABLE...
Page 52: ...a D GROUP Company 00254EN Rev 01 11 2012 52 17 WIRING DIAGRAM ASEPTIC EQUIPPED WITH CLARUS L 2...
Page 53: ...a D GROUP Company 00254EN Rev 01 11 2012 53...
Page 54: ...a D GROUP Company 00254EN Rev 01 11 2012 54...
Page 55: ...a D GROUP Company 00254EN Rev 01 11 2012 55...
Page 56: ...a D GROUP Company 00254EN Rev 01 11 2012 56...
Page 57: ...a D GROUP Company 00254EN Rev 01 11 2012 57...
Page 58: ...a D GROUP Company 00254EN Rev 01 11 2012 58...
Page 59: ...a D GROUP Company 00254EN Rev 01 11 2012 59...
Page 60: ...a D GROUP Company 00254EN Rev 01 11 2012 60 18 WIRING DIAGRAM ASEPTIC EQUIPPED WITH STERIS VHP...
Page 61: ...a D GROUP Company 00254EN Rev 01 11 2012 61...
Page 62: ...a D GROUP Company 00254EN Rev 01 11 2012 62...
Page 63: ...a D GROUP Company 00254EN Rev 01 11 2012 63...
Page 64: ...a D GROUP Company 00254EN Rev 01 11 2012 64...
Page 65: ...a D GROUP Company 00254EN Rev 01 11 2012 65...
Page 66: ...a D GROUP Company 00254EN Rev 01 11 2012 66...