SCPI Command Reference
Tunable Laser Source
111
T200S Specific Commands
[SOURce]:SWEep:HOLDoff
Description
This command sets the time duration of the pause to perform
after sweep completion.
At *RST, this value is set to 0.0 second.
Syntax
[:SOURce]:SWEep:HOLDoff<wsp><pause>[<unit>]|
MAXimum|MINimum|DEFault
Parameter(s)
³
pause
:
Numeric value corresponding to the pause in second.
To know the valid time range, use the query
on page 112 (MIN and MAX
parameters).
³
unit
Unit of the pause value: S preceded by a possible multiplier.
Possible multipliers are listed in
³
MAXimum
Sets the pause to the maximum possible value.
³
MINimum
Sets the pause to the minimum possible value.
³
DEFault
Sets the pause to the default value.
To know the default time, use the query
Example(s)
SOUR:SWE:HOLD 200 MS
SOUR:SWE:HOLD? returns 2.000000E-001
See Also