SCPI Command Reference
112
T200S
T200S Specific Commands
[SOURce]:SWEep:HOLDoff?
Description
This query returns a value indicating either the current pause
duration after sweep completion, or the
minimum/maximum/default possible pause duration.
At *RST, this value is set to 0.0 second.
Syntax
[:SOURce]:SWEep:HOLDoff?<wsp>
[MAXimum|MINimum|DEFault]
Parameter(s)
³
MAXimum
The query returns the maximum possible pause duration in
second.
³
MINimum
The query returns the minimum possible pause duration in
second.
³
DEFault
The query returns the default pause duration in second.
Response Syntax
<pause>
Response(s)
pause:
Floating point number corresponding to the current pause
duration in seconds or the minimum/maximum/default pause
duration, depending on the specified query parameters.
Example(s)
SOUR:SWE:HOLD 200 MS
SOUR:SWE:HOLD? returns 2.000000E-001