SCPI Command Reference
Variable Attenuator
115
Product-Specific Commands—Description
:INPut[1..n]:WAVelength
Description
This command selects a specific wavelength.
At *RST, the wavelength that will be selected
depends on the instrument you have.
Syntax
:INPut[1..n]:WAVelength<wsp><Wavelength[<
wsp>M]>|MAXimum|MINimum|DEFault
Parameter(s)
Wavelength:
The program data syntax for <Wavelength> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is M. The
<Wavelength> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
lowest supported value.
MAXimum allows to set the instrument to the
highest supported value.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for IQS-3150
Page 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 29: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 55: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 75: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 87: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...