SCPI Command Reference
144
IQS-3150
Product-Specific Commands—Description
a) In ABSOLUTE mode, <RelativePower> =
absolute power + power offset value.
b) In REREFENCE mode, <RelativePower> =
absolute power – power reference value +
power offset value.
c) In XB (X+B) mode, <RelativePower> =
absolute power + correction factor
(wavelength-specific) + power offset value.
Example(s)
CONT:MODE POW
OUTP:APM ABS
OUTP:RPOW -40.00 dBm
OUTP:OFFS 0.0
OUTP:RPOW? Returns -4.001
OUTP:OFFS 2.5
OUTP:RPOW? Returns -3.001
See Also
:CONTrol[1..n]MODE
:OUTPut[1..n]:APMode
:OUTPut[1..n]:OFFSet
:OUTPut[1..n]:REFerence
:OUTPut[1..n]:RPOWer
:OUTPut[1..n]:RPOWer?